background image

HGM300 Power Supply Replacement 

 

P/N 3015-5686 Rev 0 

SECTION 1:  SCOPE 

This document explains how to replace the 120 VAC power supply circuit board of an HGM300 

with a new power supply circuit board (P/N 3015-5675). 

SECTION 2:  PREREQUISITES 

Before replacing the power supply circuit board, it is important to: 

 

review all safety precautions 

 

understand specification changes (for example, new fuse specifications) 

 

familiarize yourself with the internal hardware components of the HGM300 and their 

locations within the HGM300 enclosure 

 

gather tools required to perform the replacement. 

2.1.  SAFETY PRECAUTIONS 

Be sure to review all of the safety precautions listed below before performing ANY maintenance 

on the HGM300 – including replacing the power supply circuit board. 

 

 

AC POWER SUPPLY WARNING:  

Ensure the source voltage matches the voltage of 

the product before  energizing the equipment. It is highly suggested that the 

HGM300 be placed on a separate circuit with UPS or surge protection.  A switch or 

circuit-breaker shall be included in the building installation.  It shall be in close 

proximity to the equipment and within easy reach of the operator.  It shall be 

marked as the disconnecting device for the equipment. 

 

 

PROTECTIVE GROUNDING  WARNING:  

Under no circumstances should this 

equipment be operated without connection to a protective ground. Failure to 

observe this warning poses a potential shock hazard and is also a violation of 

electrical safety standards applicable to this type of equipment.  

 

 

WORKING INSIDE THE INSTRUMENT:  

Extreme care should be exercised when 

accessing the interior of this instrument.  Only qualified electrical maintenance 

personnel should perform connections and adjustments. Always de-energize the 

power supply before working inside the instrument. 

 

 

EXPLOSIVE  ATMOSPHERE  WARNING:   

Do not operate this equipment in the 

presence of flammable liquids, vapors or aerosols. Operation of any electrical 

instrument in such an environment constitutes a safety hazard.

 

 

 

EXHAUST  VENTING  WARNING:   

It is imperative that the exhaust port on this 

instrument be properly vented as described in this manual. Failure to do so 

constitutes a safety hazard.

 

 

Summary of Contents for HGM300

Page 1: ...HGM300 Halogen Gas Monitor Power Supply Replacement Instruction 3015 5686 Rev 0 June 2012 Product Leadership Training Service Reliability...

Page 2: ...tion A switch or circuit breaker shall be included in the building installation It shall be in close proximity to the equipment and within easy reach of the operator It shall be marked as the disconne...

Page 3: ...lling into the box will void the warranty 2 2 SPECIFICATION CHANGES Table 1 summarizes the fuse requirements for old and new power supply boards For fuse locations refer to Figure 1 Figure 2 and Figur...

Page 4: ...Supply Replacement 4 P N 3015 5686 Rev 0 2 3 HARDWARE OVERVIEW Refer to the following illustrations to locate the Power Supply printed circuit board Figure 2 HGM300 Interior Components Showing Old Pow...

Page 5: ...300 Power Supply Replacement P N 3015 5686 Rev 0 5 Figure 3 Comparison of Old Style and New Style Power Supply Boards NOTE Agency approval for the new power supply has changed from UL to MET Laborator...

Page 6: ...r HGM300 It contains warnings wiring and operation information that is supplemental to this replacement procedure 2 Remove power that supplies the HGM300 Perform appropriate lock out tag out procedure...

Page 7: ...s This includes the two input power connectors the power supply connector to the pump and the two power supply connectors to the main board When installing these connectors do not use excessive force...

Page 8: ...upply Replacement 8 P N 3015 5686 Rev 0 Headquarters 621 Hunt Valley Circle New Kensington PA 15068 Website www bacharach inc com E mail help bacharach inc com Printed in U S A Registered Trademark of...

Reviews: