Fyrite INSIGHT
47
5.3 B-Smart
™
Sensor Replacement & Calibration
To replace a B-Smart
™
sensor and calibrate do the following:
1. Enter the CALIBRATION MENU per Section 5.2.
2. Use the ▲▼ buttons to select B-SMART. Press
ENTER
to display
the B-Smart code screen.
3. Use the ▲▼ buttons to enter the 10 digit alphanumeric code sup
-
plied with the B-Smart sensor. Use the ◄► buttons to move the
cursor across the screen. Press
ENTER
.
NOTE:
If the correct code was entered, the analyzer accepts it and
returns to the CALIBRATION MENU. If an incorrect
code was entered, the screen will display “Invalid Code”.
Check to make sure the correct code has been entered. If
problem persists, contact your nearest Bacharach Service
Provider.
Bacharach also offers a convenient Exchange Program that allows the
customer to regularly send in old sensors and, as scheduled, receive new
replacements that are already calibrated and that include a code that can
be entered into the analyzer for a quick, convenient setup. Contact Bacha-
rach customer service for more details about this program.
5.4 Pressure Sensor Calibration
This procedure calibrates the pressure sensor to a known pressure value.
Materials Required:
•
Bellows
•
Manometer
- Range: ± 20 mB (± 8 inwc)
- Accuracy: ± 0.025 mB (± 0.01 inwc)