Fyrite INSIGHT
7
General Characteristics:
Power Requirements
Four disposable ‘AA’ Alkaline batteries (op
-
tional AC adapter is available)
Operating Time
Minimum of 20 hours continuous operation
(pump running and backlight ON)
Warm Up Time
60 seconds
Memory
Up to 100 complete sets of
combustion, temperature, & pressure
tests can be saved in memory.
Display
128 x 64, LCD Graphic Display
Front Panel Controls
Eleven front panel push buttons
(Refer to Section 3.4)
Interfaces
IrDA printer & USB connectivity
Accuracy:
Oxygen
±0.3% O
2
with a typical flue gas
concentration of CO
2
Carbon Monoxide
±5% of reading or ±10 ppm, whichever is
greater*, in the range of 0 to 1000 ppm.
Flue Gas Temperature
±2 °C between 0 & 124 °C
(±4 °F between 32 & 255 °F)
±3 °C between 125 & 249 °C
(±6 °F between 256 & 480 °F)
±4 °C between 250 & 400 °C
(±8 °F between 481 & 752 °F)
Ambient/Primary
Temperature
±1 °C between 0 & 100 °C
(±2 °F between 32 & 212 °F)
Draft
±2% of reading or ±0.05 mB, whichever is greater
in the range of -25 to +25 mB
± 3% in the range of -25 to -100 mB
± 3% in the range of 25 to 100 mB
*
Tighter CO accuracy in the lower ranges, up to ±2 ppm, may be
attained if a lower range calibration gas (e.g. 100 ppm CO) is used.