9.0 Service
The instrument should be regularly serviced to ensure correct
operation and accurate readings. Bacharach recommends a full
service and factory calibration every 12 months.
Depending on usage, the O
2
cell should be replaced every 2-3 years.
User Serviceable Parts
NOTE:
There are no user serviceable parts inside of the instrument.
Please do not attempt any repair as this may void the
warranty.
The following parts are supplied and can be user serviced:
Sample Filter
This should be regularly inspected for damage or discoloration
and changed if needed. The instrument should never be operated
without the sample
fi
lter as this may result in water or dust entering
the instrument. The
fi
lter should be changed immediately if water
can be seen.
Sample Tubing
Always ensure that the sample tubes are not contaminated or dam-
aged.