For transmitters rated at a maximum output power not listed abo-
ve, the recommended separation distanced in meters (m) can be
estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the
higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electro-
magnetic propagation is affected by absorption and reflection from
structures, objects and people.
Guidance and manufacturer´s declarationelectromagnetic
immunity
The model e-medic
TM
Silnce TP7 is intended for use in the electro-
magnetic environment specified below. The customer or the user of
the model e-medic
TM
Silnce TP7 should assure that it is used in such
an environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic
environment
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
+ 8 kV con-
tact
+ 15 kV air
+ 8 kV con-
tact
+ 15 kV air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least 30
%.
Electrical fast
transient/burst
IEC 61000-4-4
+ 2 kV for
power
supply lines
+ 1 kV for
input/out-
put
lines
+ 2 kV for
power
supply lines
+ 1 kV for
input/output
lines
Mains power quality
should be that of a
typical commercial
or hospital environ-
ment.
Surge
IEC 61000-4-5
+ 1 kV
line(s)
to line(s)
+ 2 kV
line(s)
to earth
+ 1 kV
line(s)
to line(s)
+ 2 kV
line(s)
to earth
Mains power quality
should be that of a
typical commercial or
hospital environment.
9
Summary of Contents for e-medic SILENCE TP 7
Page 1: ......
Page 2: ...ii Version Change History Date Version Description Remark 2018 8 16 1 0 First release Cosa...
Page 30: ...17 I O parts...
Page 31: ...18 I O parts W ISO...
Page 32: ...19 I O parts W USB 3 1 gen2...
Page 35: ...22 Main...
Page 47: ......
Page 60: ...47 J16 H5 2mm J15 H9 2mm...