32
Starting a blood pressure measurement
You can cancel blood pressure measurements that are in
progress by pushing the SET button or the user 1 or user 2
touch button.
Proceed as follows to start the blood pressure monitor:
1.
On the measurement computer
, tap the user 1 touch button
or the user 2 touch button
depending on which user you
have configured for yourself in the app (see
fig. A and B
).
2.
All display elements will be shown briefly. The last measurement
will then appear on the display
for one second. If no
measurement has been performed yet, the device will not show
any values.
3.
The cuff
will now inflate and you will feel it tighten
(see
fig. A
).
4.
If no measurements have been saved, the blood pressure monitor
will inflate the cuff up to a pressure of 190 mmHg. Otherwise, the
device will calculate the average systolic pressure of all previous
measurements, add 40 mmHg and inflate the cuff to that value.
5.
As soon as the blood pressure monitor detects your pulse, the
pulse indicator
will blink on the display. If this is not the case,
the device will inflate again (see
fig. B
).
6.
Once the measurement is complete, the entire air will be released
from the cuff. The measurements for systolic pressure
,
diastolic pressure
and pulse
will appear on the display
and be saved. The LED corresponding to the measured values will
light up (see
fig. B
).
33
7.
If you do not want to perform automatic data transmission, use
the user 1 touch button or the user 2 touch button to switch the
blood pressure monitor to sleep mode (see
fig. B
).
8.
If you have prepared the blood pressure monitor and your
smartphone or tablet accordingly, automatic data transfer to the
app will begin following the measurement.
9.
The Bluetooth® symbol
will blink on the display. The blood
pressure monitor will now attempt to connect to the app for
approx. 30 seconds (see
fig. A and B
).
10.
As soon as a connection has been established, the Bluetooth®
symbol will stop blinking and all measurements will be
automatically transferred to the app. Following successful
transmission, the Bluetooth® symbol on the display will go out
and the device will switch to sleep mode. If a transmission fails,
the display will show the Bluetooth® symbol and the error code
“E7” (see chapter “Troubleshooting”).
Manual data transfer
All function buttons on the blood pressure monitor are
inactive during the data transfer.
If it is not possible to automatically transfer the measurements to your
smartphone or tablet after the measurement, you can perform a
manual transfer as follows:
1.
Activate Bluetooth® on your smartphone or tablet.
2.
Open the app.