Instructions for use
TH-134 • TH-135
En
11
8. DISINFECTION
Medical cup can be disinfected by boiling in hot water. After 3-4 minutes of boiling, shake
off excessive water from the cup and let them dry on a clean, lint free (paper) towel.
Main unit, mask, mouth piece can Not be boiled.
9. TROUBLE SHOOTING:
After checking the followings, if any problem still persists, please return the appliance to
your dealer for servicing.
Cause of failure
Causes and reaction measures
Push power
button, but
no operation
1. Check if the battery voltage is suffi cient?
When the appliance’s red indicator light is on, the battery
power is too low. Please replace the battery immediately.
2. Discard the waste batteries according to local law.
3. Check the battery is installed correctly?
Open battery cover and check if the battery electrode is rusted
or not.
Cannot
nebulize or
not enough
mist output
1. Check if the medical cup is fi lled with appropriate amount of
medicine?
2. The mesh in medical cup blocked by remaining medicine.
Cleaning medical cup, if still can’t solve problem, please replace
a new medical cup.
3. The mesh surface (mesh output) has covered by water or
medication.
Remove water or medication from mesh surface.
4. Medical cup is not installing properly.
Release the medical cup from main body and push it back again.
Suggestion:
Replace a new medical cup per year.