Instruction for Use
10
PRO-110 / PRO-115
En
the first time after long period of storage and/or if they are used by several persons).
Before using the device, proceed with the cleaning operations as described in the «9.
Cleaning, maintenance and storage» section.
•
Place the device on a stable and horizontal surface so that you can easily reach the
Nebulizer, Accessories and ON/OFF Switch when you do the inhalation.
Before using of the device make sure that:
•
all the components are properly assembled;
•
air filter is inside of the filter cap;
•
air filter is clean (if air filter has changed the color or has been used for a long period of
time, replace it with a new one).
Nebulizer kit preparation for use
1. Open the nebulizer by turning counterclockwise the top
2. Make sure that the medicine conduction cone is properly fitted on the air conduction
cone inside the nebulizer
3. Put the prescribed quantity of medicine into the nebulizer
4. Close the nebulizer by turning clockwise the two parts, paying attention that they are
thoroughly sealed
NOTICE!
Capacity of container for drugs is equal to 2-8 ml. There is a scale on the
container for drugs. It serves for rough estimate of drug solution capacity.
Air tube connection
Connect one end of the air tube to the nebulizer and
the other end to the air outlet on the device.
Make sure that air tube is properly attached in order
to avoid air leakage.
WARNING!
Hold the nebulizer strictly vertically,
do not spill the drug when attaching the air tube.
Insert the mask or mouthpiece directly onto the nebulizer
•
Make sure the switch ON/OFF switch into the «О» position.
•
Plug the device into the wall socket, making sure that
the main supply corresponds to the electrical rating
of the device.
•
Press the power-switch to the position ON «I».
7. OPERATING
Hold the nebulizer strictly vertically.
WARNING!
Do not incline the nebulizer at more than 45 degrees.
If the angle is greater than 45 degree, the drug can flow out to your mouth.
To start the treatment, set the switch into the «I» position. The compressor will start
working and a nebulizing will begin
Do the treatment in accordance with your doctor’s instructions.
Inhale the aerosol solution using the prescribed accessory.
WARNING!
The temperature of inhaled aerosol depends on the temperature of the
ambient atmosphere and the temperature of the drug solution. If the solution was
stored in refrigerator for a long period of time, it is recommended to raise the
temperature to 16-20С°.
Use of mouthpiece:
Take the mouthpiece to your mouth and breathe smoothly during the inhalation.
Using of mouthpiece is recommended in treatment of lower respiratory tract in grown-
ups and children above 5 years old.
Use of masks:
Put on a mask in the way that it will close your nose and mouth, and conduct inhalation
of the drug. Inhale and exhale through the mask.
Using of mask is recommended in treatment of upper respiratory tract. It allows for
irrigation of the whole nasal and pharynx cavities, and also larynx and trachea.
Child mask is recommended for use for children aged from 1 to 5 years old Infant mask
for children under 1 year old.
When the treatment has been completed, switch the unit off by setting the switch into
the «O» position and disconnect the plug from the wall socket.
Wash the nebulizer and its accessories as described in the section «9. Cleaning, Mainte-
nance and Storage».
Condensate can form in the air tube. In this case detach the air tube from the nebulizer,
turn on the compressor and dry the air tube till the elimination of all the liquid.
WARNING!
Do not store the air tube if there is condensate or liquid inside. It can
lead to bacterial infection.
Switch off the device from the socket.
Rinse your mouth with boiled water of the ambient temperature after the procedure. If
you used mask – rinse your eyes and face with water.
WARNING!
The device can work for 30 min without a pause, than you should let
the device to cool down for 30 min.
IM_PRO-110_PRO-115_EN_RO_BG_2916.indd 10
21.07.2016 11:39:03
Summary of Contents for PRO-110
Page 22: ...M_PRO 110_PRO 115_EN_RO_BG_2916 indd 22 21 07 2016 11 39 03...
Page 44: ...M_PRO 110_PRO 115_EN_RO_BG_2916 indd 44 21 07 2016 11 39 05...
Page 59: ...Bg PRO 110 PRO 115 59 5 A1 LOT M_PRO 110_PRO 115_EN_RO_BG_2916 indd 59 21 07 2016 11 39 07...
Page 67: ...M_PRO 110_PRO 115_EN_RO_BG_2916 indd 67 21 07 2016 11 39 08...