8 Maintenance and Cleaning
WARNING
• Before cleaning or carrying out maintenance work, always switch the unit off and disconnect the mains plug.
Working on the connected appliance may cause electric shock or fire.
• When disconnecting, always pull the plug and never the cable. Pulling the cord may result in electric shock
or fire due to a short circuit.
• During defrosting, water will leak from the unit. Collect this water so that it does not lead to a risk of slipping
for passing persons.
• Watch out that no water can penetrate the ventilation slots of the unit.
• Before cleaning or carrying out maintenance work, the contents preserved in it will have to be removed and
stored temporarily in another working refrigerator.
• Make sure the condensation water is removed on a weekly basis. Increased condensation water inside the
device may lead to malfuntioning of the device.
• Inspect the unit on a regular basis for damaged parts. Only genuine spare parts may be used. Use of any
other parts may affect the performance or cause physical injury or damage to the equipment.
CAUTION
• The use of hydrogen peroxide H202 (VHP) for disinfection (sterilization) of the appliances is not suitable.
• Other than the cleaning procedure described in this manual, follow the internal policy and procedures.
• Before using the refrigerator it should be cleaned inside and outisde.
• For cleaning, use only gentle cleaning agents. Never use aggressive or caustic cleaning agents, scouring
powder, steel, wool, abrasive sponges or chemical solvents. When cleaning, make sure that no fluids of any
kind run into the ventilation housing.
• For disinfecting, we recommend all the surface disinfecting agents commonly used by customers, provided
they are recommended by national organizations. For disinfecting small areas, we recommend using a
concentrated alcoholic agent.
8.1 Cleaning
• The device should be cleaned before initial use and at regular intervals thereafter.
• Remove any leakage immediately.
• Clean the unit (inside and outside) and dry it before any long term storage.
• The drainage vessel at the back of the unit must be emptied and cleaned regularly.
8.2 Maintenance
• The battery should be replaced every 2 years as a preventive measure.
• A defective connection cable may not be repaired. It must be replaced by customer service or by a qualified
electrical technician.
• The service personnel should be made sensitive to damages or defective parts on the equipment
immediately, so that a replacement of a defective part is possible in a reasonable period. Thus damages to
the equipment and to the load can be avoided.
Summary of Contents for MT100
Page 1: ...820 9506 71 ed0821 INSTRUCTIONS FOR USE MT100 BLOOD REFRIGERATOR INCUBATOR...
Page 2: ...2...
Page 5: ...5 1 Model Overview Group Model name Blood Refrigerator Incubator MT100...
Page 24: ...24 12 Spare parts Container door and inner equipment...
Page 25: ...25 Electronic and cooling unit...
Page 27: ...27 13 Wiring diagram 294 8062 37...
Page 28: ...B Medical Systems S r l op der Hei 17 L 9809 Hosingen Luxembourg...