7.6 Vehicle Battery Protection
• In DC-operating mode, the MT100 is equipped with a voltage management system. In order to ensure that
the vehicle battery is not completely unloaded and it can be started without problems, the MT100 is
switched off.
• The battery protection system will turn the compressor off once the supply voltage drops below a predefined
cut-off value. The compressor restarts once the battery has recovered to above the cut-off value, once the
vehicle is restarted.
• If the device is automatically switched off due to the vehicles low battery, the system generates a powerfail-
alarm (see chapter “Powerfail Alarm” for more details).
7.7 Interior Equipment
• The device is designed to have optimal function with the original baskets. If interior fittings from other
suppliers are to be used, this must be agreed with the manufacturer.
• In case the user decides to change the original fittings without prior authorization from B Medical Systems,
please assure that the air circulation inside the chamber is guaranteed in empty and loaded conditions.
Restriction of the air circulation may cause the unit not to fulfill its intended use. It is recommended that
the user validate the device with the changed configuration before operation.
Summary of Contents for MT100
Page 1: ...820 9506 71 ed0821 INSTRUCTIONS FOR USE MT100 BLOOD REFRIGERATOR INCUBATOR...
Page 2: ...2...
Page 5: ...5 1 Model Overview Group Model name Blood Refrigerator Incubator MT100...
Page 24: ...24 12 Spare parts Container door and inner equipment...
Page 25: ...25 Electronic and cooling unit...
Page 27: ...27 13 Wiring diagram 294 8062 37...
Page 28: ...B Medical Systems S r l op der Hei 17 L 9809 Hosingen Luxembourg...