SPECIFICATIONS

OUTPUT VOLTAGE O-50 VDC, continuously var-

LINE REGULATION

Maximum 0.l%, 0.32% typ-

iable. Two ranges, O-25 and

ical, at output voltage of 50

O-50 VDC. VDC and output current of 2

OUTPUT CURRENT O-2 amperes. Four ranges:

amps from 105-125 VAC.

O-50 mA, 0-0.2A, 0-0.5A, and RIPPLE

5 millivolts peak-to-peak

0-2A.

maximum.

CURRENT LIMIT 5% to 100% of each current CURRENT DERATING .02 A/°C above 25° C am-

ADJUSTMENT range.

LOAD REGULATION Maximum 0.l%, typical

SIZE

0.07%. NET WEIGHT

FEATURES

FULLY SOLID STATE Uses integrated circuits, sil-

icon transistors and diodes,

and an SCR. All the advan-

tages of solid state construc-

tion are utilized, including:

Dependability-reliability

No warm up time or stab-

ilization delay

Ruggedness

Compact size

CURRENT LIMITING/ Protects load and instrument

OVERLOAD against overload. Current

PROTECTION limit is fully adjustable from

approximately 2.5 mA to 2

A. The power supply auto-

matically shuts down and

the OVERLOAD lamp lights

if the preset current limit is

exceeded. After clearing the

cause of the overload, sim-

ply push the set/reset button

to restore normal operation.

SIMPLIFIED

CURRENT

Permits setting current limit

LIMIT SETTING

without disturbing external

load connections or output

voltage settings; does not re-

quire application of short

circuit to output terminals.

Simply push the set/reset

button and adjust the cur-

rent limit while reading the

setting on the current meter.

O-50 VDC

Continuously adjustable over

entire range with a single

control: no range switching

required.

O-2 AMPS

Divided into four ranges.

Fully regulated output at all

current levels.

DUAL METERS Allows output voltage and

current to be monitored si-

multaneously without switch-

ing ranges from current to

voltage. Both meters have

overload protection to pre-

vent damage from incorrect

range selection.

FOUR CURRENT

METER RANGES

TWO VOLTMETER

RANGES

STANDBY-DC

ON SWITCH

ON-OFF CONTROL

PILOT LAMP

MECHANICAL

PRE-REGULATOR

FLOATING OUTPUT

REVERSE POLARITY

PROTECTION

BINDING POSTS

ATTRACTIVE

APPEARANCE

EASY TO OPERATE

bient. (2 A Range only).

 x 

 x 10” deep.

11 lbs.

Provides maximum meter

resolution. Selection of me-

ter range also selects coarse

current limit setting.

0-25V and 0-50V ranges pro-

vide maximum meter resolu-

tion. A mechanical stop

prevents the voltage level

control from exceeding ap-

proximately 25 volts when

the O-25 V range is selected.

Standby mode disconnects

power supply from load

without disturbing voltage

or current control settings.

On-off switch is combined

with voltage level control to

assure voltage setting of

zero when unit is turned on.
Lights to indicate at a glance

that the unit is turned on.

Power supply’s rectifier in-

put voltage increases in pro-

gressive steps as output

voltage setting is increased.

Improves efficiency; less

power is converted to heat

at low voltage operation.

Permits referencing the posi-

tive or negative output to

any external dc potential or

ground.
Protects against accidental

damage from reverse polar-

ity connection to external

power source.
Heavy duty S-way binding

posts for positive and nega-

tive polarity output and earth

ground.
Modem, functional design.

When more than one unit is

used, units may be stacked.

All controls are identified

and easy to read. Simplified

operation helps prevent op

erator mistakes that might

damage equipment.

3

Summary of Contents for 1601

Page 1: ...Solid State Regulated DC Power Supply A Product of DYNASCAN CORPORATION l 6460 West Cortland Street l Chicago Illinois 60635 PRICE 3 66 ...

Page 2: ...rvicing Vehicular Electronics Equipment 7 Servicing Plug In Modules 9 Using Two Power Supplies For Two Output Voltages 9 TWO Power Supplies In Series For O 100 Volt Output 9 Using The Power Supply As A Battery Charger 10 Other Servicing Applications 10 Electronics Manufacturing 10 Electronics Design Engineering 11 Electronics Education 11 CIRCUIT DESCRIPTION 12 MAINTENANCE 13 FUSE REPLACEMENT 14 ...

Page 3: ...y regulated output at all current levels DUAL METERS Allows output voltage and current to be monitored si multaneously without switch ing ranges from current to voltage Both meters have overload protection to pre vent damage from incorrect range selection FOUR CURRENT METER RANGES TWO VOLTMETER RANGES STANDBY DC ON SWITCH ON OFF CONTROL PILOT LAMP MECHANICAL PRE REGULATOR FLOATING OUTPUT REVERSE P...

Page 4: ...xercises in basic and advanced electronics 1 current meter 2 1 l3 l 2 Controls and Indicators Measures actual output load 4 Push to set current during normal oper CURRENT ation LIMIT or to reset Indicates current limit value OVERLOAD when set reset button 4 is button set pushed reset button 2 CURRENT Selects range for current RANGE switch meter 1 and coarse setting of current limit Full scale mete...

Page 5: ...nput of the device being powered to the terminal 10 of the power supply C If the positive polarity of the device being powered is also to be ground reference jumper the terminal 9 to the ter minal 8 If the negative polarity of the device being powered is to be ground ref erence jumper the terminal 10 to the terminal 8 If neither the positive nor negative polarity of the device being pow ered is to...

Page 6: ......

Page 7: ......

Page 8: ......

Page 9: ......

Page 10: ...B output level is pre regulated in coarse steps As the LEVEL control is rotated clockwise from zero to maximum the unregulated B voltage changes from its lowest to its highest value in four steps This minimizes the difference between the unregulated B and the regulated B output which always keeps power dissipation with in safe limits The main components which make up this circuit are power transfo...

Page 11: ......

Page 12: ......

Page 13: ......

Page 14: ......

Reviews: