CAUTION:
Do not fill less than minimum or exceed maximum fill volume (see below).
TABLE 1
– Delivery Time Information
When filled to nominal volume, flow rate accuracy is ±15% of the labeled infusion rate
when infusion is started 0-8 hrs. after fill and delivering nominal saline as the diluent at
88˚F (31˚C) against a back pressure of 40 cm of water.
CAUTION:
Filling the pump less than nominal, increases flow rate. Filling the pump
greater than nominal, decreases flow rate.
Easypump RA
270-5
270-10
300-6
400-5
400-8
400-10
Nominal Flow Rate (ml/hr)
5
10
6
5
8
10
Nominal Fill Volume (ml)
270
270
300
400
400
400
Maximum Fill Volume (ml)
335
335
335
550
550
550
Retained Volume (ml)
≤
8
≤
8
≤
8
≤
15
≤
15
≤
15
Time to Reach Room Temperature
hours from refrigerator
12
12
12
12
12
12
hours from freezer
24
24
24
24
24
24
Approximate Delivery Time
Fill Volume (ml)
24 h
1 d
150
250
170
230
280
30 h
1.3 d
290
36 h
1.5 d
325
235
225
300
370
48 h
2 d
255
285
270
400
460
60 h
2.5 d
290
335
310
72 h
3 d
330
350
96 h
4 d
440
120 h
5 d
135 h
5.5 d
6 d