Infusomat® fmS 2.3 gb
8 - 1
8
8 -
Procedural Instructions on the TSC
Visual Inspection
1.
Check the Infusomat® fmS and accessories for cleanliness.
2.
Check the Infusomat® fmS and accessories for completeness
and check configuration. Pay special attention to the follow-
ing parts:
a)
Check the special additional labels of the Infusomat®
fmS adjusted for the Infusomat® Space Line, please see
3.
Check the Infusomat® fmS and its accessories for damage
and the labels for readability. Pay special attention to the fol-
lowing parts:
a)
Pump sealing diaphragm
b)
Membrane keyboard
c)
Rubber feet
d)
Operating unit
e)
Lock mechanism of the pump cover
f)
Flow inhibitor
g)
Mains plug connector
h)
MFC plug connector
i)
Check fuse values
A fuse with T 0.315 A must be installed for devices for
100/110/120 V.
A fuse with T 0.16 A must be installed for devices for
200/230/240 V.
Fig.: 8 - 1
8 - 8
Summary of Contents for 871 5416
Page 1: ...Service Manual Infusomat fmS 0 Version 2 3 English 0 10 ...
Page 10: ...Contact Persons 0 0 10 Infusomat fmS 2 3 gb For your notes ...
Page 22: ...System Overview 1 1 12 Infusomat fmS 2 3 gb For your notes ...
Page 30: ...Software 2 2 8 Infusomat fmS 2 3 gb For your notes ...
Page 60: ...Unit Elements 4 4 20 Infusomat fmS 2 3 gb For your notes ...
Page 64: ...Maintenance 6 6 2 Infusomat fmS 2 3 gb For your notes ...
Page 78: ...Test Equipment and Special Tools 9 9 2 Infusomat fmS 2 3 gb For your notes ...
Page 82: ...Spare Parts List 10 10 4 Infusomat fmS 2 3 gb For your notes ...
Page 84: ...Index 11 11 2 Infusomat fmS 2 3 gb ...
Page 86: ...Revision Documentation A A 2 Infusomat fmS 2 3 gb For your notes ...