Unit Elements
4
Infusomat® fmS 2.3 gb
4 - 13
-
Air alarm: air bubbles in ml and air rate in ml/h respectively
-
User language
-
Special functions (ON/OFF)
-
Menu
-
Staff call type
Note
If data is not entered, ”Calibration data faulty” may be displayed
after the unit is switched on again.
9.
After ending the Service Program save the data.
The counters for operating hours, battery operation and pump
head cycles are reset to zero when the pump unit is exchanged. If
”Calibration data faulty” is displayed, select the Service Program
and quit with ”Save? Yes”.
Check
Safety check, pump unit check.
Summary of Contents for 871 5416
Page 1: ...Service Manual Infusomat fmS 0 Version 2 3 English 0 10 ...
Page 10: ...Contact Persons 0 0 10 Infusomat fmS 2 3 gb For your notes ...
Page 22: ...System Overview 1 1 12 Infusomat fmS 2 3 gb For your notes ...
Page 30: ...Software 2 2 8 Infusomat fmS 2 3 gb For your notes ...
Page 60: ...Unit Elements 4 4 20 Infusomat fmS 2 3 gb For your notes ...
Page 64: ...Maintenance 6 6 2 Infusomat fmS 2 3 gb For your notes ...
Page 78: ...Test Equipment and Special Tools 9 9 2 Infusomat fmS 2 3 gb For your notes ...
Page 82: ...Spare Parts List 10 10 4 Infusomat fmS 2 3 gb For your notes ...
Page 84: ...Index 11 11 2 Infusomat fmS 2 3 gb ...
Page 86: ...Revision Documentation A A 2 Infusomat fmS 2 3 gb For your notes ...