Checks after Repair
5
5 - 2
Infusomat® fmS 2.3 gb
Observe the procedure information
Pressure limitation, mechanical
(Reference values measured with
electronic pressure sensor)
❒
With original Infusomat® line
❒
P
max
❒
P
min
❒
With Infusomat® Space line
❒
P
max
❒
Value P
max
by min. 0.10 bar higher
than pressure level, high, electronic
❒
P
min
Safety clamp (flow inhibitor)
❒
P
min
Delivery accuracy
Ambient temperature 20 ... 25 °C
Delivery rate: 250 ml/h
Measured volume: 25 ml
❒
Divergence
Drop sensor
Delivery rate: 400 ml/h
❒
Drop sensor alarm, occlusion
❒
Drop sensor alarm, flow
Air inline sensor
Delivery rate: 250 ml/h
❒
Water value
❒
Air alarm
❒
Air value
❒
Threshold value
Battery test
Operation mains/battery/mains:
❒
Uninterrupted operation
Switch on unit without mains connection
❒
Self-test is executed
1. Visual Inspection
2. Electrical Safety
according to IEC / EN 60601-1
or VDE 0750 and VDE 0751
3. Functional Inspection
(Part 2 of 2)
Summary of Contents for 871 5416
Page 1: ...Service Manual Infusomat fmS 0 Version 2 3 English 0 10 ...
Page 10: ...Contact Persons 0 0 10 Infusomat fmS 2 3 gb For your notes ...
Page 22: ...System Overview 1 1 12 Infusomat fmS 2 3 gb For your notes ...
Page 30: ...Software 2 2 8 Infusomat fmS 2 3 gb For your notes ...
Page 60: ...Unit Elements 4 4 20 Infusomat fmS 2 3 gb For your notes ...
Page 64: ...Maintenance 6 6 2 Infusomat fmS 2 3 gb For your notes ...
Page 78: ...Test Equipment and Special Tools 9 9 2 Infusomat fmS 2 3 gb For your notes ...
Page 82: ...Spare Parts List 10 10 4 Infusomat fmS 2 3 gb For your notes ...
Page 84: ...Index 11 11 2 Infusomat fmS 2 3 gb ...
Page 86: ...Revision Documentation A A 2 Infusomat fmS 2 3 gb For your notes ...