This Service Manual was approved by B. Braun on 16.03.2006.
This manual has been completely revised. The most important
changes are listed below:
-
Changed manual structure
-
New software
-
New spare parts
-
Total list of spare parts
-
Checking the pressure cut-off during Bolus after repair
Current Information
If you hear a scraping noise when the drive arm is pulled out, the
straight pin (under the spindle) may have come loose. In this case,
an additional straight pin lock (Ord. No. 3450 9100) can be insert-
ed in units up to serial No. 66725. From serial No. 66725 on this
straight pin lock is already fitted. Observe the instructions at-
tached.
A - 2
Summary of Contents for 0871 4827
Page 1: ...Service Manual Perfusor compact Version 2 1 english 0 0 10 ...
Page 10: ...Contact Persons 0 0 10 Perfusor compact 2 1 gb For your notes ...
Page 14: ...System Overview 1 1 4 Perfusor compact 2 1 gb For your notes ...
Page 20: ...Software 2 2 6 Perfusor compact 2 1 gb For your notes ...
Page 54: ...Unit Elements 4 4 18 Perfusor compact 2 1 gb For your notes ...
Page 60: ...Checks after Repair 5 5 6 Perfusor compact 2 1 gb For your notes ...
Page 62: ...Maintenance 6 6 2 Perfusor compact 2 1 gb For your notes ...
Page 69: ...Procedural Instructions on the TSC 8 8 5 Perfusor compact 2 1 gb ...
Page 70: ...Procedural Instructions on the TSC 8 8 6 Perfusor compact 2 1 gb For your notes ...
Page 72: ...Test Equipment and Special Tools 9 9 2 Perfusor compact 2 1 gb For your notes ...
Page 76: ...10 4 Perfusor compact 2 1 gb For your notes ...
Page 80: ...Appendix A A A 2 Perfusor compact 2 1 gb For your notes ...
Page 82: ...Modification Instructions for Syringe Holder B B 2 Perfusor compact 2 1 gb For your notes ...