Service Program
3
Perfusor® compact, 2.1 gb
3 - 3
Configuration
1.
Select the language, interface and the screen display desired
in the
File
➨
Configuration
menu.
2.
Acknowledge with
OK
.
Working with the Service Program
Preparation
1.
Connect service cable (
) to MFC connector
) and the PC serial port (COM 1 or COM 2).
2.
Connect mains cable to the unit.
3.
Start the Service Program on the PC.
4.
To start communication press the ON key on the Perfusor®
compact until „Release On/Off key“ is displayed on the PC.
and
are displayed on the LC-display.
Display / Save the Unit Settings
1.
Menu
File
➨
Connect
.
2.
Menu
EEPROM
➨
Read
.
3.
Menu
File
➨
Save
.
4.
Call menu
Modes
➨
Modification
and menu
Syringes
➨
Sy-
ringe selection
or
Syringe types
. Note down parameters pri-
or to any modification (e.g. new E-Module).
Adjust Unit Settings
1.
Menu
File
➨
Connect
.
2.
Menu
EEPROM
➨
Read
.
3.
Desired modifications / display, please see:
-
Operation
➨
Operation values
-
Modes
➨
Modification data
-
Calibration
➨
Pressure calibration
(required in case of
bolus rate change)
-
Syringes
➨
Syringe selection
or
Syringe types
-
Constants
➨
Service interval
4.
Menu
EEPROM
➨
Write
transmits data to the device. Menu
File
➨
Save
saves the data on the hard disc.
5.
Enter the user number 0 upon query.
6.
Check unit according to the procedura instructions for in-
spection
(see „Procedural Instructions for Inspection after
Modifications via the Service Program“
Fig.: 3 - 2
Legende zu Abb. 3 - 2:
ItemDesignation
1
MFC connector on the unit
2
MFC service cable
Summary of Contents for 0871 4827
Page 1: ...Service Manual Perfusor compact Version 2 1 english 0 0 10 ...
Page 10: ...Contact Persons 0 0 10 Perfusor compact 2 1 gb For your notes ...
Page 14: ...System Overview 1 1 4 Perfusor compact 2 1 gb For your notes ...
Page 20: ...Software 2 2 6 Perfusor compact 2 1 gb For your notes ...
Page 54: ...Unit Elements 4 4 18 Perfusor compact 2 1 gb For your notes ...
Page 60: ...Checks after Repair 5 5 6 Perfusor compact 2 1 gb For your notes ...
Page 62: ...Maintenance 6 6 2 Perfusor compact 2 1 gb For your notes ...
Page 69: ...Procedural Instructions on the TSC 8 8 5 Perfusor compact 2 1 gb ...
Page 70: ...Procedural Instructions on the TSC 8 8 6 Perfusor compact 2 1 gb For your notes ...
Page 72: ...Test Equipment and Special Tools 9 9 2 Perfusor compact 2 1 gb For your notes ...
Page 76: ...10 4 Perfusor compact 2 1 gb For your notes ...
Page 80: ...Appendix A A A 2 Perfusor compact 2 1 gb For your notes ...
Page 82: ...Modification Instructions for Syringe Holder B B 2 Perfusor compact 2 1 gb For your notes ...