Unit Elements
4
4 - 16
Perfusor® compact, 2.1 gb
3.
Unscrew four screws.
4.
Remove cover for drive head housing with toggle, lever and
release shaft (square).
5.
Sketch the cable layout.
Note
Pay attention to spring when removing the housing cover.
6.
Remove cam switch with pressure spring and pressure pin.
7.
Remove push-button sensor board and disconnect plug con-
nector.
8.
Push drive head housing towards the unit.
9.
Pull jaws off the drive tube.
10. Pull drive head housing together with clamp off the drive
tube.
11. Assembly is done in reverse order.
Note
Pay attention to cable layout. The cable must be laid between the
pins and under the light barrier.
Fig.: 4 - 20
Drive head housing
Cap for drive head housing
Release shaft
Toggle
Pressure spring
Lever
Cam switch
Sensor board
Fig.: 4 - 21
Jaws
Drive tube
Axial positioner
Clamp
Drive head housing
Seal washer
Fig.: 4 - 22
Connection cable of push-button
sensor
Summary of Contents for 0871 4827
Page 1: ...Service Manual Perfusor compact Version 2 1 english 0 0 10 ...
Page 10: ...Contact Persons 0 0 10 Perfusor compact 2 1 gb For your notes ...
Page 14: ...System Overview 1 1 4 Perfusor compact 2 1 gb For your notes ...
Page 20: ...Software 2 2 6 Perfusor compact 2 1 gb For your notes ...
Page 54: ...Unit Elements 4 4 18 Perfusor compact 2 1 gb For your notes ...
Page 60: ...Checks after Repair 5 5 6 Perfusor compact 2 1 gb For your notes ...
Page 62: ...Maintenance 6 6 2 Perfusor compact 2 1 gb For your notes ...
Page 69: ...Procedural Instructions on the TSC 8 8 5 Perfusor compact 2 1 gb ...
Page 70: ...Procedural Instructions on the TSC 8 8 6 Perfusor compact 2 1 gb For your notes ...
Page 72: ...Test Equipment and Special Tools 9 9 2 Perfusor compact 2 1 gb For your notes ...
Page 76: ...10 4 Perfusor compact 2 1 gb For your notes ...
Page 80: ...Appendix A A A 2 Perfusor compact 2 1 gb For your notes ...
Page 82: ...Modification Instructions for Syringe Holder B B 2 Perfusor compact 2 1 gb For your notes ...