background image

OPERATION MANUAL

Technical changes reserved

0141 0316-25   08.02.2021

USB-I²C-adapter

for B+B Temperature-module with I²C-output

Connection layout of plug module

Quick start with ‚TERATERM‘

TERATERM is an universal, efficient terminal program which you can find in the 

installationversion on the Download-Center. Download-Center:  https://down-

load.bb-sensors.com/index.php/login → Login Name: bb-shop, Password: BB

-

shop#2017. TERATERM is a free ware; you can find the license requirements 

and copyright note in the installation directory.
Connect the sensor module to the USB adapter. If necessary, use the adapter 
cable supplied with the adapter.
Now connect the USB adapter to the PC. The installation of the driver should 

automatically  take  place  in  Windows.  If  driver  files  are  required,  these  are 

available on the ZIP in the directory „Driver“ for different operating systems.

In rare cases, when the installation of hardware does not work at the first att

-

empt, you can intervene in the control panel. Further hints on this can be found 
in the FAQs on the CD. After th hardware is operational, the message „The new 
hardware can now be used“ appears on the screen.
Now install the terminal program TERATERM from the enclosed CD and start 
the program. Select the interface under „Serial port“ where the device is con-
nected and set the following communication parameters:

Then carry out the following settings under „Terminal settings“:

After this, the system should be operational.

Reading the values

Note: <CR> means “pressing the carriage return key”.
First test the connection with the hardware: For each character sent, the red 

LED on the USBAdapter shortly blinks. If the device answers a
telegram, the green LED also blinks. First send the character V <CR>. In res

-

ponse, the USB adapter should reply back the version string:

As per the revision status of the software the contents can vary, if needed, but 
a response must come.
If you can not establish the connection here, then check the hardware in Win-
dows control panel and also all other settings mentioned before.After the con-

nection to the module is functional, first you must initialise the adapter:

T11200 <CR>

To start the reading, the operating voltage to the sensor must be switched on:

IRT78004 <CR>

After this command, the LED near the probe connection glows red and it shows 

that the operating voltage is present at the measuring probe.The sensor replies 

with 8 characters, in the example it is :3F3944C1. These are the first set of 

measured values.

6-pole multi-pin connector

1

VDD

Supply voltage 6...12 V

2

GND

Ground

3

SDA

Serial Data I²C

4

SCL

Serial Clock I²C

5

VRH/OWI

RH Voltage Output/OWI

6

---

Not used

B+B Thermo-Technik GmbH | Heinrich-Hertz-Straße 4 | D-78166 Donaueschingen 

Fon +49 771 83160 | Fax +49 771 831650 | [email protected] | bb-sensors.com

2 / 4

Summary of Contents for USB-ADA-12C

Page 1: ...th I C output Simple serial ASCII protocol Operation possible over terminal Areas of application Probe testing product development Interface for own PC based products Configuring I C Bus address Calibration of ASICs Technical data PC USB adapter for I C Interface USB interface 1 1 and 2 0 compatible Dimensions B x H x T 79 x 21 x 39 mm Sensor connection Over RJ11 RJ12 plug 6 pin Serial communicati...

Page 2: ...e is con nected and set the following communication parameters Then carry out the following settings under Terminal settings After this the system should be operational Reading the values Note CR means pressing the carriage return key First test the connection with the hardware For each character sent the red LED on the USBAdapter shortly blinks If the device answers a telegram the green LED also ...

Page 3: ...n point Pressure module Pressure un calibrated Command overview for examples The USB adapter has a very extensive command set Detailed description is available on the enclosed CD The following explanations are only related to the commands used in the above examples V Statement of version string This command responds back with the version string of the controller Firm ware T11200 Initialising Defin...

Page 4: ... 300 C THMOD I2C 300 Measuring range 270 800 C THMOD I2C 800 Measuring range 270 1370 C THMOD I2C 1370 Temperature module Pt1000 with voltage output 0 5 V and I C bus Article Article number Measuring range 32 95 9961 C TEMOD I2C R1 Measuring range 32 223 992 C TEMOD I2C R2 Measuring range 32 479 984 C TEMOD I2C R3 B B Thermo Technik GmbH Heinrich Hertz Straße 4 D 78166 Donaueschingen Fon 49 771 83...

Reviews: