7
External load
ID 4 mm × Length 3 m + 20 cc or over
Air supply capacity
20 L/minute(N) or over, with 6.7 kPa {50 mmHg} change
Air consumption
5 L/minute(N) or less (when balanced at output 100%)
Accuracy
KDP11; ±0.5% F.S. (For span 50 to 500 kPa {5000 to 50000 mmH
2
O})
±0.75% F.S. (For span 25 to less than 50 kPa {2500 to less than 5000 mmH
2
O})
KDP22; ±0.5% F.S. (For span 5 to 53.9 kPa {500 to 5500 mmH
2
O})
±0.75% F.S. (For span 2.5 to less than 5 kPa {250 to less than 500 mmH
2
O})
KDP33; ±0.5% F.S. (For span 1.0 to 6 kPa {100 to 600 mmH
2
O})
±1.0% F.S. (For span 0.5 to less than 1 kPa {50 to less than 100 mmH
2
O})
KDP44; ±0.5% F.S. (For span 0.2 to 1.2 kPa {20 to 120 mmH
2
O})
±1.0% F.S. (For span 0.1 to less than 0.2 kPa {10 to less than 20 mmH
2
O})
KDP61/71/81;
±0.5% F.S. (For span 50 to 500 kPa {5000 to 50000 mmH
2
O})
±1.0% F.S. (For span 25 to less than 50 kPa {2500 to less than 5000
mmH
2
O})
KDP62/72/82;
±0.5% F.S. (For span 5 to 53.9 kPa {500 to 5500 mmH
2
O})
±1.0% F.S. (For span 2.5 to less than 5 kPa {250 to less than 500 mmH
2
O})
Dead band
0.1% F.S.
Damping (KDP44)
Time constant: Minimum 2 sec. or less, Maximum 25 sec. or over, continuously variable.