Instrument maintenance and troubleshooting
Azbil Corporation
4-6
Model JTD720A - Smart Multivariable Transmitter
(5) Wiring and Piping During Calibration
In general, the following wiring and piping should be employed.
+
-
CommStaff
V
P
DC Power
Supply 24 V
Voltmeter
Precision Resistor
Red Wire
Black Wire
Standard
Pressure Generator
Connects to pressure
inlet of cover flange
Figure 4-3 Wiring and piping during Calibration
(6) Checking the Setting Range
See the operation manuals for the respective communicators.
Field Communication Software CommStaff Model: CFS100 Instruction Manual
(Smart Transmitter Model JTD720A Edition)
CM2-CFS100-2007
(7) Measurement Range Calibration Procedure
See the operation manuals for the respective communicators.
Field Communication Software CommStaff Model: CFS100 Instruction Manual
(Smart Transmitter Model JTD720A Edition)
CM2-CFS100-2007
•