ME-02 USER MANUAL MD004
39
_____________________________________________________________________________________________
Press Program key.
Press Enter key when MEtEr is displayed, then
choose CALIb.
Following options will display:
-CAL on – full load calibration (Max),
-CAL StP – full load calibration with confirmation of
successive steps F,
out – out without calibration
Press Enter key when CAL StP is displayed
(calibration in two steps confirmed by Enter).
Press Enter key when standard of mass value is
displayed, which will be used for calibration or use
othEr option and insert proper value.
Press Enter and wait until scale zero is stored,
signalised by „- - - - - - ”
After LOAd sign appears put on standard of mass
and press Enter key.
Wait for the calibration end and weight displaying
.
Attention:
Using option CAL on instead of CAL StP
uwalnia frees from double pressing Enter key.
0
0
MODE
MODE
g
g
g
MODE
MODE
MODE
1000g