Internal calibration options:
Internal calibration of the balance is performed automatically every time the balance is
switched on, additionally after given time interval during work and after every
temperature change of more than given value.
In order to perform internal calibration in any moment, empty the pan and press key
twice (one more pressing terminates calibration).
Press
MENU
key to display function
menu and choose
CALIb
function by
pressing
T
key when it is displayed.
The following options will appear:
-
CAL on
–
perform calibration with
external weight
-
CAL Prn
–
printout of calibration report
-
CAL tM
–
set time interval for internal
calibration (1h
–
6h)
-
CAL
C
- set temperature difference for
internal calibration (1
C - 4
C)
-
out
–
switch internal calibration off
for internal calibration
Press
T
key when
CAL tM
option is
displayed
.
Predefined time intervals for
internal calibration will be displayed.
Select required value pressing
T
key
.
Accordingly choose
CAL
C
option
pressing
T
key and selecting values
of temperature difference.
Select
out
option to finish.
M O DE
M O DE
M O DE
M O DE
M O DE
M O DE
M O DE