01
Channel
2 channels – with adjustable intensity
02
Intensity
Adjustable, 0 – 100 mA (in steps of 1 mA)
with 500 ohm load on each channel
03
Pulse amplitude
Asymmetrical, Bi-phasic square pulse
04
Output voltage
0 – 50 V
05
Power source
1 pc. 9V battery
06
Size
10.2cm (L) x 6.2cm (B) x 2.8cm (H)
07
Weight
150 grams (batteries included)
08
Pulse
rate
Adjustable, from 1 – 150 Hz, 1 Hz/step
09
Pulse
width
Adjustable, from 50 – 300 microseconds,
10 μ/step
10
ON
time
Adjustable, from 2 – 90 seconds, 1 Sec./step
11
OFF
time
Adjustable, from 0 – 90 seconds, 1 Sec./step
12
Ramp
time
Adjustable, from 1 – 8 seconds, 1 Sec./step
13
Mode
5 TENS modes: B (Burst) N (Normal),
M (Modulation), SD1 (40% modulation),
SD2 (70% modulation) – 2 EMS modes: S
(Synchronous), A (Alternate) – P (Pro-
gram Mode) – 24 TENS programs (set by
default)
14
TENS Burst mode
B
Burst rate: Adjustable, 0.5 – 5 Hz – Pulse
Width adjustable, 50 – 300 μs –
Fixed internal frequency = 100 Hz
15
TENS Normal mode
N
The pulse rate and pulse width are adjus-
table. It generates continuous stimulation
based on the setting value
16
TENS Modulation
mode
M
Modulation mode is a combination of pul-
se rate and pulse width modulation. The
pulse rate and width are automatically
varied in a cycle pattern. The pulse width
is decreased by 50% from its original
setting in 0.5 second, then the pulse rate
is decreased by 50% from its original
setting in 0.5 second. Total cycle time is
1second. In this mode, pulse rate (1 – 150
Hz) and pulse width (50 – 300 μs) are
fully adjustable.
TECHNICAL DESCRIPTION
46
AXION
EN
Summary of Contents for STIM-PRO X700
Page 19: ...Vorderseite R ckseite BER DAS GER T 18 AXION DE...
Page 43: ...Front Back ABOUT THE DEVICE 42 AXION EN...
Page 67: ...Face avant Face arri re A PROPOS DE L APPAREIL 66 AXION FR...
Page 91: ...Parte delantera Parte trasera DISPOSITIVO 90 AXION ES...
Page 115: ...Davanti Dietro SUL DISPOSITIVO 114 AXION IT...
Page 139: ...PRZ D URZ DZENIA TY URZ DZENIA O URZ DZENIU 138 AXION PL...