Set pulse width
The pulse width can be
adjusted from 50 µs to 300
µs. It regulates the time for
which the electrical signal is
applied. The current value is
shown on the LCD display.
Timer
The device comes with a tim-
er that can be adjusted from
1 to 60 minutes. It can also
be set to continuous mode.
The application ends
automatically when the timer
runs out.
Set ramp time
This setting is only available in EMS mode.
The ramp time controls the
time of output current that
increase from 0 to the setting
value, and from the setting
value back to 0. The ramp
time is adjustable from 1 to 8
seconds.
Set contraction time
(ON-Time)
This setting is only available in EMS mode.
The ON time controls the
contraction time. By pressing
the
control, the contrac-
tion time is adjustable from
2 – 90 seconds. This time in-
40
AXION
EN