7
To Recap:
1. Select Channel
2. Select Preset
3. Press START
Adjustable Parameters
The micro5 provides you with complete control over the following parameters; microamps, frequency,
waveslope, polarity and time. This allows for maximal flexibility when programming treatment param
-
eters for each patient
The Parameter Ranges Are As Follows:
Manual Override
To enter settings other than the presets proceed as follows:
Micro-amps
Select the channel(s) to be programmed. When the micro-amp field is flashing you may enter any range from
1-750uA using the numerical key pad. If the micro-amp field is the only field you want to change you do not
need to press the ENTER key, you may proceed to the START key. For example if you wanted to enter 600 uA
you would press 6,0,0.To enter 60uA you do not need to press 0,6,0. Simply press 6,0. If you make an entry
error or change your mind you can press the CLR key to clear the entry and start again. Pressing CLR clears
only the parameter being programmed and not the remaining parameters. NOTE: any entry above 750uA will
default back to 750uA. After the settings are entered you can move to the next parameter by pressing the ENT
key. To pass over a parameter simply keep pressing the ENT key until you have reached the parameter you
want to program.
Frequency
When entering the frequency settings use the 0 key to enter frequencies less than one. For example to enter
.6Hz you would press 0,6. To enter numbers greater than .9 simply enter the number direct. l5OHz would be
enter as 1,5,0. If you want to change the waveslope press ENT after the frequency has been entered.