ONE.

 

1.  

Low Battery. If batteries are too low, proper output cannot be maintained. If the battery charge indicator

 

 is flashing, charge the batteries as soon as possible. See section on batteries to determine voltage.

2. 

Defective lead wires. If there is a break in the lead wires INSTANT ALERT will show a current test fail   

 

ure. To test for lead wire failure plug a set of lead wires into channel 1. Press the PROBE button to enter

  

into the probe mode. Touch the exposed lead wire metal tips together. The LCD probe display should

            read over 400. If it does not this means a break somewhere in the lead wire. Shake the lead wire to see  

            if the break causes an intermittent problem. Replace the defective lead wire.

3. 

Defective electrodes. Carbon Silicone electrodes have a natural useful life of about 6-8 months. Bad  

 

 

electrodes can impede the flow of current and cause INSTANT ALERT to issue a current test failure. Try

  

a new set of electrodes.

Multi-application electrodes have several components that can break down and cause a failure. Even new elec-

trodes can be bad. Try another pair.

CAUTION: THERE ARE NO USER SERVICEABLE PARTS OTHER THAN BATTERIES. DO NOT OPEN 

THE INSTRUMENT WITHOUT INSTRUCTIONS FROM THE COMPANY.

The probe is also connected to the INSTANT ALERT system. Breaking contact with the skin during treatment 

will cause a current test failure.

Note: INSTANT ALERT current test operates by capturing samples to measure against what is entered on the 

screen. At lower frequencies it may take the microprocessor a little extra time to find a sample to measure. This 

will account for the slight lag in showing a failure in the lower frequencies.

 

The AXION micro5 uses a combination of power sources.

Display Memory

One button type lithium battery not accessible to the user. When this battery fails the instrument will no longer 

display the last entry upon start up. The unit must be returned to us to replace this battery.

Main Processor Board and Output

Three high capacity “D” size NiMH batteries. Recharge these batteries using the supplied USB charger, Com-

puter USB port or auto USB port. Charging time is about 6 hours for a full charge. When charging make sure 

the charging indicator light is on. This light is located on the back of the micro5. The low battery indicator will 

turn on when it is time to recharge. THE UNIT CANNOT BE USED DURING CHARGING.

To keep the channels isolated, channel two has its own rechargeable power source. These batteries are recharged 

at the same time the main batteries are recharged so no extra steps are needed.

During charging the micro5 will turn it self off. This prevents operation with the charger plugged in. TURN 

THE  MICRO5 OFF BEFORE CHARGING. Should the device ever fail to turn on, make sure that the charging 

cord is not plugged into the back of the device. If unit still fails to turn on, contact Company.

Summary of Contents for micro5

Page 1: ...ESSOR CONTROLS ARE DESIGNED TO SIMPLIFY TREATMENT DELIVERY WE HAVE MADE THE OPERATION OF THE micro5 VERY SIMPLE IF YOU FOLLOW THE BASIC GUIDE LINES OF THE INSTRUCTION MANUAL REMEMBER MOST PROBLEMS ARE...

Page 2: ...ssociated with high output milliamperes instruments Microcurrent stimulation is usually sub sensory and often imperceptible to the patient The current is able to pass through the skin with out much st...

Page 3: ...essing this green key sets the parameters of the selected channels for the E default 4 M Pressing this yellow key sets the parameters of the selected channels for the M default 5 I Pressing this white...

Page 4: ...the screen 12 START Pressing this key starts the countdown operation of the instrument It will also silence the beep that is heard every ten seconds during operation 13 STOP Pressing this key suspends...

Page 5: ...ost effective applications available for microcur rent stimulation The following table illustrates the parameter setting for each of the modes These values will appear on the LCD screen when the mode...

Page 6: ...when a large area is treated This preset uses both channels to deliver the current through four pads The micro5 uses both a rate scan and vector scan When using the I preset you will notice the numbe...

Page 7: ...example if you wanted to enter 600 uA you would press 6 0 0 To enter 60uA you do not need to press 0 6 0 Simply press 6 0 If you make an entry error or change your mind you can press the CLR key to cl...

Page 8: ...CLR to make your change You may program both channels at once by pressing the COMB combined key You may press the START key at any time during programming You do not have to be in the time setting to...

Page 9: ...y is a difference Never use the plastic shaft they don t conduct electricity and be sure to watch the instruction video on the proper insertion of the Q Tip Reusable electrodes have a life span of abo...

Page 10: ...metal sides of the probe See Video 4 Scan the area to be treated You will notice that when contact with the skin is made the micro5 emits a tone This tone rises and falls in pitch as the resistance i...

Page 11: ...ecial salts compound It s clear that when using moistened q tips they should be soaked in a saline solution Doing that eliminates one variable you can control 1 2 Teaspoon of our mi cro5 Electrolyte s...

Page 12: ...w tones indicates a successful initial start up test During Operation A series of descending tones along with a flashing battery indicator and flashing test indicator means a low battery charge You sh...

Page 13: ...e a current test failure Note INSTANT ALERT current test operates by capturing samples to measure against what is entered on the screen At lower frequencies it may take the microprocessor a little ext...

Page 14: ...or pain of central origin 3 T E N S devices should be used only under the continued supervision of a licensed practitioner 4 T E N S devices have no curative value 5 T E N S is a symptomatic treatment...

Page 15: ...n in the continuous mode until NEXT 1 is pressed again or until the instrument is turned off by the On Off switch on the back of the unit Display Current NEXT 2 This feature is used to display the rel...

Page 16: ...Sequential Treatment chaining The micro5 allows the clinician to link two treatments together to form a single continuous treatment This is sometimes called chaining To sequence treatments 1 Press the...

Page 17: ...the I preset you will notice that both chan nels are ready for programming BOTH CHANNELS MUST BE USED The microamp default for both chan nel is 400 Static Full Field with vector scan The output of ch...

Page 18: ...e the charging module in the wall when not in use Leaving it in the wall will shorten the life of the charging module The micro5 cannot be used while it is charging IMPORTANT NOTE It takes the micro5...

Page 19: ...AXION micro5 We have tested and recommend the Anker Astro3E Our tests show that the Anker can recharge the micro5 from dead batteries to 90 charge in 1 hour NOTE It does not power the micro5 It recha...

Page 20: ...el 1 The second powers Channel 2 The D cell battery pack is located inside the unit and is acessible by removing the four screws around the handle on the back of the unit and carefully sliding the tra...

Page 21: ...ants the AXION micro5 to be free from defects for a period of two years from date of pur chase Factory further warrants all accessories including batteries and chargers to be free from defects for a p...

Reviews: