Phase Angle (PG=acos (PF))
Range
Accuracy
Resolution
Measuring Condition
0
O
~90
O
(capacitive or inductive)
±2
O
1
O
The minimum measuring current
10A
The minimum measuring voltage
45V
0
O
~90
O
(capacitive or inductive)
For reference only
Measuring current less than 10A
OR
Measuring voltage less than 45V
Remarks:
Allowable maximum overload protection voltage: 750V RMS
Allowable maximum overload protection current: 1000A RMS
Basic Functions
Range
Best Accuracy
AC Voltage
100V/400V/750V
±(1.2%+5digits)
AC Current
40A/100A/400A/1000A
±(2%+5 digits)
Active Power
0.01kW-750kW
±(3%+5 digits)
Apparent Power
0.01kVA-750kVA
±(3%+5 digits)
Reactive Power
0.01kVAr-750kVAr
±(4%+5 digits)