PEE1V03
20-11- 2017
5
2. TECHNICAL DATA
Accuracy class
2 by LST EN1434-1:2007.
Energy units
kWh, MWh, GJ, Gcal
Maximum value of thermal power
5,28 MW
Flow measurement
The ratio of the permanent flow rate to the lower limit of the flow-rate
(the user selects during order):
q
p
/q
i
100 ,
or q
p
/q
i
250 (only for flow sensors with q
p
1,5 m
3
/h; 2,5 m
3
/h; 6,0 m
3
/h; 15 m
3
/h)
Flow sensor can be delivered for threaded connection (up to q
p
=10,0 m
3
/h) or flanged.
Technical data of the flow sensor are presented in table 1.1
Table 1.1
Permanent
flow rate qp,
m
3
/h
Upper
flow rate q
s
,
m
3
/h
Lower
flow rate qi,
m
3
/h
Threshold
value of flow
rate, m
3
/h
Overall
length L,
mm
Pressure
losses at
q
p
, kPa
Joining to the pipeline
(Thread
– G,
flange
–DN)
0,6
1,2
0,006
0,003
110
7
G3/4“
0,6
1,2
0,006
0,003
190
0,9
G1“ arba DN20
1
2
0,01
0,005
110
11,3
G3/4“
1
2
0,01
0,005
190
2,5
G1“ arba DN20
1,5
3
0,006
0,003
110
17,1
G3/4“
1,5
3
0,006
0,003
190
5,8
G1“ arba DN20
1,5
3
0,015
0,003
110
17,1
G3/4“
1,5
3
0,015
0,003
190
5,8
G1“ arba DN20
1,5
3
0,015
0,005
130
7,2
G1“
2,5
5
0,01
0,005
130
19,8
G1“
2,5
5
0,01
0,005
190
9,4
G1“ arba DN20
2,5
5
0,025
0,005
130
19,8
G1“
2,5
5
0,025
0,005
190
9,4
G1“ arba DN20
3,5
7
0,035
0,017
260
4
G1 1/4“arba DN25
6
12
0,024
0,012
260
10
G1 1/4“arba DN25
6
12
0,06
0,012
260
10
G1 1/4“arba DN25
10
20
0,04
0,02
300
18
G2“ arba DN40
10
20
0,100
0,02
300
18
G2“ arba DN40
15
30
0,06
0,03
270
12
DN50
15
30
0,15
0,03
270
12
DN50
25
50
0,1
0,05
300
20
DN65
25
50
0,25
0,05
300
20
DN65
40
80
0,16
0,08
300
18
DN80
40
80
0,4
0,08
300
18
DN80
60
120
0,24
0,12
360
18
DN100
60
120
0,6
0,12
360
18
DN100
Temperature limits of heat conveying liquid:
5
o
C....130
o
C.
Note:
For heat conveying liquid temperature below 90
o
C, the calculator can remain on the flow sensor or
be mounted on the wall.
For heat conveying liquid temperature above 90
o
C
the calculator must be mounted on the
wall.
Connection cable length between the calculator
and the flow sensor
1,2 m
(2,5 m or 5,0 m- according to the special order)
Maximum admissible working pressure
16 bar or 25 bar.
Behavior of the meter, when the flow rate exceeds the maximum value q
s
:
- At the flow rate q
< 1,2 ∙ q
s
-linear,
- At the flow rate q
> 1,2 ∙ q
s
- constant (
q = 1,2 ∙ qs is applied to calculations of thermal energy).
The error
„Maximum allowable value of flow rate is exceeded“ is registered and duration of error is calculated.
Pulse inputs (additional)
Number of pulse inputs
2
Measurement
unit’s
m
3
Pulse value
programmable