Sensitivity: Virtual value 1.5V. Overload protection: 250V DC or AC peak value. Duty cycle display:(0.1%-
99.9%)
8.2.7. AC/DC Measurement
• Turn the function/range SELECT key to “2000A” or higher range as shown in the right figure.
• Press the SELECT key to choose AC or DC measurement mode.
• If used for devices nearby electromagnetic field, unstable or incorrect readings may be displayed.
• Please press the REL key to reset before current measurement.
• Press the head trigger to open the head and use the head to clamp the conductor to be tested and then
release the trigger slowly until the head is closed completely. Please confirm if the conductor is
clamped in the center of the head, otherwise it will cause additional error. This meter can only
measure 1 current conductor at a time; if it measures two or more current conductors at the same
time, the measurement reading will be incorrect.
DC Technical Indicators:
AC Technical Indicators:
1 - Range
2 - Accuracy
3- Resolution
Note