Thank you for purchasing the American Weigh 

ZEO-50 portable milligram scale. Please read all operating 

instructions carefully before use. Scales are precision 

measuring devices and should always be handled with 

proper care. To ensure years of reliable service, keep these 

simple tips in mind:

• 

Do not exceed the scale’s maximum capacity. 

Overloading your scale can permanently damage it!

• 

Avoid exposure to extreme heat or cold. Scales 

perform best at normal room temperature. If 

temperatures have changed dramatically, calibration 

and adjustment may be necessary.

• 

Allow your scale to warm up for 30-60 seconds before 

performing calibration and adjustment(if available).

• 

Store your scale in a clean, dry location. Dust, dirt, 

and moisture can accumulate on the weighing 

sensors and electronics causing inaccuracy or 

malfunction.

• 

Avoid using your scale in close proximity to strong 

radio wave emitting devices such as computers, cash 

registers, and cordless phones.

• 

Always weigh on a flat and level surface, free from 

vibrations and drafts. The corner of a room is usually 

the most stable.

• 

Gently apply all items to be weighed. Do not drop 

items onto the weighing platform.

• 

Avoid dropping your scale. The warranty does not 

cover damage due to rough treatment or overload.

• 

Check the batteries first if you are having any trouble 

with your scale. This simple step can remedy most 

scale issues.

• 

Do not disassemble your scale. This product contains 

no user serviceable parts.

Summary of Contents for ZEO-50

Page 1: ...ZEO 50 50g x 0 001g User Manual ZEO 50 Copyright 2013 American Weigh Scales Inc All rights reserved Rev 1 0...

Page 2: ...calibration and adjustment if available Store your scale in a clean dry location Dust dirt and moisture can accumulate on the weighing sensors and electronics causing inaccuracy or malfunction Avoid...

Page 3: ...container 4 PCS Key Press to enter counting mode Hold to set the sample quantity size III Battery Installation 1 Before using the scale for the first time check the battery compartment and remove any...

Page 4: ...ure 2 If the scale is turned off and then back on it will default to the last unit used ZEO 50 Units g gram oz ounce ozt troy ounce dwt pennyweight ct carat gn grain VI Using the Tare Feature You can...

Page 5: ...nge the sample size to match the number you placed on the scale 4 Press the PCS key to confirm the sample size 5 The display will now show pcs next to the value You may remove the sample quantity and...

Page 6: ...splay will flash 50 000g 4 Place the 50g calibration weight on the center of the platform Once the weight is accepted the display will show PASS then return to normal weighing Calibration is complete...

Page 7: ...ded Remove the extra weight to avoid damaging the load cell 3 L Batteries are low Replace the batteries ZEO 50 Max Capacity 50g Readability 0 001g Power 4 x AAA Dimensions 5 9 x 3 8 x 2 5 Platform Dim...

Page 8: ...ZEO 50 American Weigh Scales Inc Norcross GA 30092 Made In China to our exact specifications and quality control...