User Manual AWR Nebulizer MN-300/2
Page 12 of 32 (Version MN-300/2-EN-V3)
4.
Intended Use
The AWR Nebulizer MN-300/2 is a nebulizer designed to introduce exact doses of liquids
in the form of aerosol into the airways of the lungs.
WARNING
The AWR Nebulizer MN-300/2 is not an officially certified medicine product and
therefore not approved for nebulizing medications. Usage and choice of possible
medications for nebulizing can only be determined by the treating physician. The
chosen medications have to be approved by the manufacturers (pharmaceutical
companies) for use with this type of application (inhalation via ultrasonic
nebulizing).
Step 1: Filling the nebulizer unit:
The liquid recommended for inhalative lung care is a 0.9 % saline solution.
-
Open the sealing cap by turning it 80 degrees
counter-clockwise and then lifting it up.
-
Fill the liquid into the chamber.
-
Re-attach the sealing cap and close it by turning
it 80 degrees clockwise.
NOTE
The initial setup for the AWR Nebulizer MN-300/2 is being described here for use
with the nebulizer unit with an MMD of 4.0 µm. This information, however, is
also valid for the use with the optionally available nebulizer unit with an MMD of
6.0 µm.