1.Connect a cable from the SYNC OUT connector to the TRIG input of a sampling
oscilloscope. Connect the IN port of the output module to the OUT port on the rear
panel of the mainframe using the supplied coaxial cable. Attach the laser diode
anode to the output module “OUT” terminal. Connect a 711S current probe (or
equivalent) as shown below. Connect the current probe to the oscilloscope. Set the
oscilloscope to trigger externally.
2.Turn on the AVO-6A1-B. The main menu will appear on the LCD.
3.To set the AVO-6A1-B to trigger from the internal clock at a PRF of 10 kHz:
a)The arrow pointer should be pointing at the frequency menu item. If it is not, press
the MOVE button until it is.
b)Press the CHANGE button. The frequency submenu will appear. Rotate the
ADJUST knob until the frequency is set at 10 kHz.
c)The arrow pointer should be pointing at the “Internal” choice. If it is not, press
MOVE until it is.
d)Press CHANGE to return to the main menu.
4.To set the delay to 1 us:
a)Press the MOVE button until the arrow pointer is pointing at the delay menu item.
b)Press the CHANGE button. The delay submenu will appear. Rotate the ADJUST
knob until the delay is set at 1 us.
c)The arrow pointer should be pointing at the “Normal” choice. If it is not, press
MOVE until it is.
d)Press CHANGE to return to the main menu.
5.To set the pulse width to 1 us:
27
OUT
CONNECTOR
SYNC
OUTPUT
AC
POWER
SAMPLING
OSCILLOSCOPE,
BW > 5 GHz
50 OHM INPUT
TRIG
INPUT
AVO-6
MAINFRAME
GND
AVO-6A1-T
GND
Bold lines = coaxial cables
20
dB
ATTENUATOR
ATTENUATOR
DUT
CURRENT
PROBE