Since the AVO-9H-B-P1B can generate pulses with subnanosecond rise times, it may
be necessary to use a sampling oscilloscope, rather than a real-time oscilloscope. In
this case, the test arrangement should be altered as shown below:
Since most sampling oscilloscopes have limited input amplitude ranges, attenuators are
required. Select appropriate attenuators for your oscilloscope.
When the output module is not used, a 50 Ohm load impedance is
required
for proper
test operation.
BASIC TEST ARRANGEMENT - WITH OUTPUT MODULE
To fully test the instrument, and for normal operation, the output module should be
connected as shown below:
21
AVO-6
MAINFRAME
SMA OUTPUT
CONNECTOR
SYNC
OUTPUT
AC
POWER
ALL CABLES: 50 OHM COAXIAL
REAL-TIME
OSCILLOSCOPE
CHANNEL A
TRIG
INPUT
SCOPE
PROBE
50Ω
TEST
LOAD
AVO-6
MAINFRAME
SMA OUTPUT
CONNECTOR
SYNC
CONNECTOR
AC
POWER
ALL CABLES: 50 OHM COAXIAL
SAMPLING
OSCILLOSCOPE
BW > 5 GHz
50 OHM INPUT
TRIG
CONNECTOR
60 dB
ATTENUATOR