![Avtech AV-1030-B Instructions Manual Download Page 30](http://html1.mh-extra.com/html/avtech/av-1030-b/av-1030-b_instructions-manual_3034527030.webp)
In the above example, the pulse width of the output was set by a programming
command. To set the output pulse width to track the trigger pulse width in external
mode, use:
*rst
(resets the instrument)
trigger:source external
(selects internal triggering)
pulse:width in
(PW
OUT
= PW
IN
)
output:type TTL
(sets the logic outputs to TTL mode)
volt:ampl 5
(sets the amplitude to 5 V)
output on
(turns on the output)
These commands will satisfy 90% of your programming needs.
ALL PROGRAMMING COMMANDS
For more advanced programmers, a complete list of the available commands is given
below. These commands are described in detail in the “Programming Manual for -B
Instruments”. (Note: this manual also includes some commands that are not
implemented in this instrument. They can be ignored.)
Keyword
Parameter
Notes
DIAGnostic:
:AMPLitude
:CALibration
<numeric value>
[no query form]
LOCAL
OUTPut:
:[STATe]
<boolean value>
:PROTection
:TRIPped?
[query only]
:TYPE
TTL | ECL
REMOTE
[SOURce]:
:FREQuency
[:CW | FIXed]
<numeric value>
[SOURce]:
:PULSe
:PERiod
<numeric value>
:WIDTh
<numeric value>
:DCYCle
<numeric value>
:HOLD
WIDTh | DCYCle
:DELay
<numeric value>
:DOUBle
[:STATE]
<boolean value>
:DELay
<numeric value>
:GATE
:TYPE
ASYNC | SYNC
:LEVel
HIgh | LOw
[SOURce]:
:VOLTage
[:LEVel]
30
Summary of Contents for AV-1030-B
Page 42: ...PERFORMANCE CHECK SHEET 42...