a)Press the MOVE button until the arrow pointer is pointing at the output menu item.
b)Press the CHANGE button. The output submenu will appear.
c)Press MOVE until the arrow pointer is pointing at the “ON” choice.
d)Press CHANGE to return to the main menu.
8.To change the output amplitude:
a)Press the MOVE button until the arrow pointer is pointing at the amplitude menu
item.
b)Press the CHANGE button. The amplitude submenu will appear. Rotate the
ADJUST knob until the amplitude is set at +5V.
c)Observe the oscilloscope. You should see 50 ns wide, 5V pulses. If you do not,
you may need to adjust the delay setting to a value more compatible with your
sampling oscilloscope. Repeat step 4 if required. You may also need to adjust the
sampling scope controls.
d)Rotate the ADJUST knob. The amplitude as seen on the oscilloscope should vary.
Return it to 5V.
e)Press the +/- button on the front panel. The amplitude as seen on the oscilloscope
should flip polarity, to -5V.
f) Press CHANGE to return to the main menu.
This completes the operational check.
28