OPERATIONAL CHECK
This section describes a sequence to confirm the basic operation of the instrument. It
should be performed after receiving the instrument. It is a useful learning exercise as
well.
Before proceeding with this procedure, finish reading this instruction manual
thoroughly. Then read the “Local Control” section of the “Programming Manual for -B
Instruments” thoroughly. The “Local Control” section describes the front panel controls
used in this operational check - in particular, the MOVE, CHANGE, and ADJUST
controls.
BASIC TEST ARRANGEMENT
1. Connect the pulse generator to a sampling oscilloscope as shown above. Note that:
a) The use of 60 dB attenuator at the sampling scope vertical input channel will
insure a peak input signal to the sampling scope of less than 1 Volt. WARNING:
This model may provide a peak output power in excess of 6 kW. The peak
power rating of the attenuator must exceed this limit. Factory tests are
conducted using Midwest Microwave model ATT-0527-20-SMA-07 attenuators.
b) The TRIG output channel provides TTL level signals (approximately 0 and
+3V). To avoid overdriving the TRIG input channel of some scopes, a 20 dB
attenuator should be placed at the input to the scope trigger channel.
c) The bandwidth capability of components and instruments used to display the
pulse generator output signal (attenuators, cables, connectors, etc.) should
exceed 1 GHz.
AVTECH
PULSER
MAIN OUTPUT
CONNECTOR
TRIG
CONNECTOR
AC
POWER
ALL CABLES: 50 OHM COAXIAL
SAMPLING
OSCILLOSCOPE
BW > 1 GHz
50 OHM INPUT
TRIG
CONNECTOR
60 dB
ATTENUATOR
26
Summary of Contents for AVRK-3-B
Page 36: ...PCB 158R4 LOW VOLTAGE POWER SUPPLY...
Page 37: ...PCB 284B HIGH VOLTAGE POWER SUPPLY...
Page 38: ...PCB 216E OP AMP TWEAKS...
Page 43: ...PERFORMANCE CHECKSHEET 43...