![AVTECH ELECTROSYSTEMS AVG-4A-B Instructions Manual Download Page 21](http://html1.mh-extra.com/html/avtech-electrosystems/avg-4a-b/avg-4a-b_instructions-manual_3034693021.webp)
will complete the full pulse width if the output is high, and then stop triggering. Pulses
are not truncated.
AMPLITUDE DRIFT
After setting a particular amplitude, it may take up to 3 minutes (typically) for the pulse
generator circuitry to reach thermal equilibrium. The initial amplitude may be up to
~10% less than the set amplitude until the circuitry has warmed up. For this reason,
amplitudes should be verified by oscilloscope measurements in applications where
accuracy important.
21
Summary of Contents for AVG-4A-B
Page 34: ...PCB 158R5 LOW VOLTAGE POWER SUPPLY...
Page 35: ...PCB 168C HIGH VOLTAGE DC POWER SUPPLY...
Page 36: ...PCB 104G KEYPAD DISPLAY BOARD...
Page 38: ...PERFORMANCE CHECK SHEET 38...