EMC Tables
Table 2 For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration-electromagnetic immunity
The
NT16B TOUCH FREE Infrared Thermometer
is intended for use in the
electromagnetic environment specified below. The customer or the user of
the
NT16B TOUCH FREE Infrared Thermometer
should assure that it is
used in such an environment.
Immunity
test
IEC 60601
Test level
Compliance
Level
Electromagnetic
environment-guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6kV contact
±8kV air
±6kV
contact
±8kV air
Floor should be wood,
concrete or ceramic tile. If
floors are covered with
shythetic material, the
relative humidity should
be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV air for
input/output
lines
Not
applicable
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
±1kV differential
mode
±2kV common
mode
Not
applicable
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines
IEC 61000-4-11
<5% UT (>95%
dip in UT for 0.5
cycle
40% UT (60%
dip in UT for 5
cycles
70% UT (30%
dip in UT for 25
cycles
5% UT (>95%
dip in UT for 5
sec
Not
applicable
Mains power quality
should be that of a typical
commercial or hospital
environment. If the use of
the TS41
Ear/Forehead/Object
Thermometer requires
continued operation
during power mains
interruptions, it is
recommended that the
TS41
Ear/Forehead/Object
Thermometer be powered
from an uninterruptible
power supply or a battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3A/m
Power frequency
magnetic fields should be
at levels characteristic of
a typical location in a
typical commercial or
hospital environment.
NOTE
is the a.c. mains voltage prior to application of the test level.
17
EMC Tables
Table 3 For EQUIPMETN and SYSTEMS that are not
LIFE-SUPPORTING
Guidance and manufacturer’s declaration-electromagnetic immunity
The
NT16B TOUCH FREE Infrared Thermometer
is intended for use in the
electromagnetic environment specified below. The customer or the user of
the
NT16B TOUCH FREE Infrared Thermometer
should assure that it is
used in such an environment.
Immunity
test
IEC 60601
Test level
Compliance
Level
Electromagnetic
environment-guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3Vrms
150kHz to 80
MHz
3V/m
80MHz to 2.5
GHz
3 Vrms
3 V/m
Portable and mobile RF
communications equipment
should be used no closer to
any part of the
NT16B
TOUCH FREE Infrared
Thermometer,
including
cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance.
d=1.2
√
P
d=1.2
√
P 80 MHz to 800 MHz
d=2.3
√
P 800 MHz to 2.5 GHz
where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site
survey,” should be less than
the compliance level in each
frequency range.” Interference
may occur in the vicinity of
equipment marked with the
following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
Field strengths from fixed transmitters, such as base stations from radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast can not be predicted theoretic
call with accuracy. To assess the electromagnetic environment due to fixed
RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the
NT16B TOUCH FREE
Infrared Thermometer
is used exceeds the applicable RF compliance level
above, the
NT16B TOUCH FREE Infrared Thermometer
should be
observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the
NT16B TOUCH FREE Infrared Thermometer
.
Over the frequency range 150kHz to 80MHz, field strengths should be less
than [V1] V/m.
18