18
To calibrate your WI-150, you must enter the Configuration Menu outlined
below. If you are already in the Configuration Menu, go directly to the
procedures for setting Zero & Span and Linearity and viewing Display which
are continued on the next page.
1.
While in Gross/Net Weighing Mode, scroll in the security code number
150 by using the
↑
and
←
keys.
2.
With the number
150 displayed, press and hold the MENU key until SET
UP is displayed.
DO NOT let go of the MENU key until
SET UP is displayed or else TARE will
be displayed. If this occurs, press the GROSS/NET key to return to Weigh-
ing Mode and begin again at Step 1.
3.
Press MENU to display
ADJUST.
4.
Press SELECT to display
ZERO.
5.
You are now in the Configuration Menu and may calibrate your system.
To move around within the Configuration Menu follow the instructions
printed in the box below. Specific instructions for setting Zero & Span
and Linearity and viewing Display are provided on the next page.
Calibration Procedures
Make sure your test weights
match the selected unit of
measure on your indicator.
To Enter the
Configuration Mode:
Summary of Contents for WI-150
Page 1: ...WI 150 Ultra Low Power Weight Indicator Service Manual ...
Page 9: ...6 Figure 2 Configuration Menu ...
Page 10: ...7 Key Available only when SC 150 is enabled Default values ...
Page 27: ...24 ...
Page 29: ...WI 150 INDICATOR AC DC BARRIER POWERED CABLE TO PC BOARD CONNECTION and CABLE PIN OUTS 26 ...
Page 30: ...WI 150 AC DC INDICATOR BATTERY POWERED CABLE TO PC BOARD CONNECTION 27 ...
Page 31: ...WI 150 INDICATOR SYSTEM BLOCK DIAGRAM 28 ...
Page 38: ...SC 150 REMOTE CONTROL OPTION BOARD MAIN BOARD 35 ...
Page 41: ...SC 150 REMOTE CONTROL SYSTEM BLOCK DIAGRAM 38 ...