Calibration
You will only be allowed to calibrate the scale when using full service access.
1. Unplug the scale from the power supply.
2. Break the tamper seal, carefully remove the blanking plate and plug the service
tool into the side of the scale.
3. Replace the weighplate and re-connect the power supply.
4. Check that the gravity factors are correct (see page 13).
5. Place a full load on the scale and remove it several times in order to 'exercise
the scale'.
6. Enter calibration mode:
7. Make sure there is no load on the scale.
8. Place a half load on the scale.
9. Place a full load on the scale.
10. Remove half the load.
20
Service Instructions
Calibration
CAL
0.0005
C
CAL 1
0.0005
C
CAL 2
10.5476
C
CAL 3
19.8754
C
CAL 4
10.5476
C