Model 6700 Family User’s Manual
6
Under Capacity Limits
Over Capacity Limits
Sealing
Internal Counts
Dynamic Response
Communications
Under capacity indication will be given with
dashes appearing on the bottom line of the
display whenever the display is below the initial
zero value.
Over capacity indication will be given with
dashes appearing in the upper line of the
display whenever the weighed item exceeds 9
divisions over the rated capacity of the unit. The
scale will use the initial zero value for reference
for over capacity determination.
Access to the calibration switch can be secured
with a lead wire or pressure sensitive security
seal. The remote or primary indicators have no
metrological features that require the use of a
security seal.
The scale has 100,000 internal counts.
The time interval when weight is applied to the
scale until a stable weight is displayed:
0–1000d,1.5 seconds
1000d+, 2.0 seconds
maximum mean average
Factory default settings: 9600 baud, 7 data bits,
even parity, 1 stop bit.
Specific cable requirements are determined by
particular POS terminal. A standard 9-pin pass
through RS-232 interface cable is necessary for
most PC interfaces. Not a null modem.
RS-232 bidirectional, configurable 1200 to
19.2K baud. Transmits weight and scale status
whenever ASCII “W” <CR> is sent by the POS
terminal. Standard ECR protocol is
OPOS
compatible. For Protocol details contact factory.