Model 6700 Family User’s Manual
13
The Diagnostic
(DIAG)
mode menu allows
testing of specific areas of the scale’s function
and viewing current configuration settings. This
includes the following:
DISPLAY (DISP)
– Shows the version and
revision of the software, followed by a display
segment test.
RAM (RA)
– Performs a nondestructive test of
RAM in the processor.
ROM (RO)
– Performs a checksum of all loca-
tions of ROM in the processor.
INPUT/OUTPUT (I/O)
– Data is output by the
scale and through the use of a loopback con-
nector. The data is immediately read back into
the receive channel and verified against what
was sent. Requires a jumper (short) between
transmit and receive data lines.
DIVISION TEST, w/AZT (DIV-A)
– Weight data
is normalized to 100,000 counts of displayed
resolution. AZT (Auto Zero Tracking) is enabled.
DIVISION TEST, w/o AZT (DIV-N)
– Weight
data is normalized to 100,000 counts of dis-
played resolution. AZT is disabled.
Viewable configuration settings include:
LN FR
Line frequency
FILT
Filter setting
PROT
Serial protocol
BAUD
Baud rate
CAP
Scale capacity
UNITS
Unit of measure
Diagnostics Mode
Diagnostic (DIAG) Mode
Tip: Easy access to
the diagnostic mode is
available directly from
the front panel without
opening the scale to
access the Menu
mode switches. See
Setup Review/Test
section.