background image

 

 

ID

020393

 

 

Revision: 

1-00

 

© Avantes BV 

www.avantes.com

 

[email protected] 

 

CC connection 

Position to insert CC-VIS/NIR of the system to be calibrated. CC can be locked with an allen screw. 
 

I/O connector 

For automatic disable control to be controlled by software it is needed to connect an AvaSpec 
spectrometer to the AvaLight-HAL-CAL-Mini with the interface cable IC-DB26-2.  
 
The signals on the I/O connector are described below. 
 

Pin configuration 
Pin  Name 

 

Description 

Vout 

Output 

Nominal 12VDC output 

10 

GND 

 

Ground 

11 

Disable 

Input 

Input, used to disable/shut down the output 

 
Below the electrical characteristics for the input signals. 
 

Electrical characteristics 

Name 

Low level detection  High level detection  Maximum voltage 

Disable 

< 1.0V 

> 2.0V 

13.2V 

 

2.2

 

12VDC Power plug 

 
The power supply specifications are mentioned in the table below 

Name 

Minimum  

Typical  

Maximum  

Voltage 

10V 

12V 

15V 

Power  

18Watt 

 

 

 

2.3

 

Settings for disable 

 
With the disable function that is present on the I/O connector the user can disable the optical 
output of the light source by setting this signal to high-level. 
 
When the light source is used in combination with an AvaSpec spectrometer, interface cable IC-
DB26-AvaSphere-0.5 and AvaSoft, the output can be disabled or enabled by AvaSoft.  
See appendix A. 
 

Summary of Contents for AvaLight-Hal-CAL-Mini

Page 1: ......

Page 2: ...oduct specifications and descriptions in this manual are subject to change without notice Avantes BV will not be responsible for errors and omissions in this manual or for direct or incidental damages...

Page 3: ...roduction 3 2 AvaLight HAL CAL Mini 4 2 1 Quick start 4 2 2 12VDC Power plug 5 2 3 Settings for disable 5 3 AvaLight HAL CAL ISP Mini 6 3 1 Quick start 6 3 2 Settings for disable 8 4 Technical support...

Page 4: ...halogen light source The new AvaLight HAL CAL Mini has a new feature called the Disable function With the Disable function the bulb is turned off to safe power and lifetime Changing of the bulb Since...

Page 5: ...AL Mini 2 1 Quick start Power LED The green LED act as status LED for the light source when the power is turned on ON OFF Switch Choose operating mode with the ON OFF switch Switch position Light sour...

Page 6: ...used to disable shut down the output Below the electrical characteristics for the input signals Electrical characteristics Name Low level detection High level detection Maximum voltage Disable 1 0V 2...

Page 7: ...ID 020393 Revision 1 00 6 Avantes BV www avantes com info avantes com 3 AvaLight HAL CAL ISP Mini 3 1 Quick start Sphere Coupler ON OFF switch Power LED 12VDC Power plug I O connector...

Page 8: ...isable control to be controlled by software it is needed to connect an AvaSpec spectrometer to the AvaLight HAL CAL Mini with the interface cable IC DB26 2 The signals on the I O connector are describ...

Page 9: ...rdered unit The bottom plate will hold the AvaSphere into position 3 2 Settings for disable With the disable function that is present on the I O connector the user can disable the optical output of th...

Page 10: ...ests concerning your AvaLight HAL CAL MINI or the AvaSoft software please go to http www avantes com In case you have any questions or need support please send us an email via support avantes com 4 2...

Page 11: ...ion ISP80 bottom plate 258 x 130 x 35 mm optional extended range NIR calibration 5 1 Ordering Information AvaLight HAL CAL Mini NIST traceable Halogen Lamp with CC VIS NIR diffuser including PS 12V 2...