3

 

Replacement of Deuterium or Halogen Lamp 

 

Warning

:  

 

Make sure the Deuterium Lamp is switched OFF before disconnecting the power cord.  

Before re-connecting the power cord, make sure that the power switch, the Deuterium lamp and 

the Halogen lamp switch are switched off.  

 

Warning:

  

 

During operation the lamp envelope reaches a temperature of approx. 250°C. Should you intend to 

change the lamp please wait for at least 20 minutes for cooling before touching the lamp. The lamp 

envelope (quartz glass) should not be touched with bare fingers, otherwise lifetime will be severely 

reduced. 

 

Please follow these instructions for the bulb replacement: 

 

1.

 

Switch off the light source and disconnect the unit from the mains power supply. 

2.

 

Remove both side covers by unclicking. 

3.

 

Remove the 4 screws securing the top cover by using a torx screwdriver to remove the top 

cover. 

4.

 

Disconnect the deuterium power connector. 

5.

 

Dismount the front panel by unscrewing the 4 front screws. 

6.

 

Carefully slide out the front panel until the green connector is visible and disconnect the 

connector of the switches. 

 

 

 

7.

 

Turn the light source upside down and carefully slide out the front panel. 

 

Switch 

connector 

Deuterium 

connector 

Summary of Contents for AvaLight-D-S

Page 1: ......

Page 2: ...oduct specifications and descriptions in this manual are subject to change without notice Avantes BV will not be responsible for errors and omissions in this manual or for direct or incidental damages...

Page 3: ...Shutter Open 5 2 5 Power connector with double Fuse and switch 5 2 6 Halogen light power adjustment 6 2 7 I O connector 6 3 Replacement of Deuterium or Halogen Lamp 7 4 Technical support 9 4 1 Getting...

Page 4: ...DUV version starts even lower at 175 nm for your deep UV experiments This version also offers twice the intensity at 200 nm The output of the AvaLight DH S is optimized for fibers or bundles upto 600...

Page 5: ...the SMA plug of the Light Source Connect your Fiber optic Cable first before starting the Deuterium Lamp Please be sure to avoid direct radiation of skin and eyes at the fiber output TTL shutter Shut...

Page 6: ...tching the Halogen switch downwards to ON the Halogen lamp switches on The Halogen LED indicates the operation mode Switching the Halogen switch upwards to OFF the Halogen lamp switches off Switch pos...

Page 7: ...he halogen power 2 7 I O connector For automatic shutter control to be controlled by software it is needed to connect an AvaSpec spectrometer to the AvaLight D H S with the interface cable IC DB26 2 N...

Page 8: ...he lamp envelope quartz glass should not be touched with bare fingers otherwise lifetime will be severely reduced Please follow these instructions for the bulb replacement 1 Switch off the light sourc...

Page 9: ...crews to secure the deuterium lamp 12 Remove the 2 screws of the halogen lamp fitting 13 Take out the halogen lamp fitting 14 Replace the halogen bulb with new 5W bulb caution do not touch the lamp gl...

Page 10: ...ches are ON before main power is switched on Put buttons for Deuterium and Halogen in OFF position before switching on main power 4 3 2 Deuterium Lamp Only FAULT POSSIBLE CAUSE REMEDY Power switch on...

Page 11: ...gen Lamp Only FAULT POSSIBLE CAUSE REMEDY Halogen Lamp does not work after switching on Halogen Lamp defective Replace Halogen Lamp LED does not light up after switching on Halogen lamp Internal Power...

Page 12: ...0 mm ca 5 kg total power for the specified wavelength range 5 1 Ordering Information AvaLight D S Deuterium light source 190 400 nm incl TTL shutter SR fibers needed AvaLight DH S Deuterium Halogen li...