2.2

 

Deuterium Light 

 

By switching the Deuterium switch downwards to ON the Deuterium lamp is preheated. After a 

warm-up time of 10-20 sec. the Deuterium lamp will ignite automatically. 

After successful ignition the Deuterium LED lights up green and indicates that the Deuterium lamp is 

in operation. Should the Deuterium lamp not ignite the Deuterium LED lights up red. This indicates a 

malfunction of the lamp. See section “4.3 Trouble Shooting”.

 

Switch to OFF to reset the Error LED. 

Switching the Deuterium switch to upwards to OFF the Deuterium lamp switches off. 

 

Switch position 

Deuterium light 

Up 

OFF 

Down 

ON (after 10-20 seconds) 

 

2.3

 

Halogen Light 

 

Switching the Halogen switch downwards to ON, the Halogen lamp switches on. The Halogen LED 

indicates the operation mode. 

Switching the Halogen switch upwards to OFF, the Halogen lamp switches off. 

 

Switch position 

Halogen light 

Up 

OFF 

Down 

ON 

 

2.4

 

TTL Shutter – Open 

 
This switch is used to open and close the shutter either manually or by TTL pulse. 
The LED indicated if the shutter open (LED on) or closed. 
 

Position switch 

Shutter 

Up 

Closed or TTL activated 

Down 

Open 

 

2.5

 

Power connector with double Fuse and switch 

 
The power connector for 100-240 VAC is located on the rear of the AvaLight-D(H)-S. Be careful to 
use it with a designated power range only, please use the included power cord with the instrument. 
For UK, US and Australian power cords, contact Avantes Technical Support.  
The 2 Fuses are 1A slow blowing Fuses. 
 

 

 

Disconnect power before opening housing or replace Fuse. 

The installation

 

category for this equipment is Class 2. 

 

 
 

 

Summary of Contents for AvaLight-D-S

Page 1: ......

Page 2: ...oduct specifications and descriptions in this manual are subject to change without notice Avantes BV will not be responsible for errors and omissions in this manual or for direct or incidental damages...

Page 3: ...Shutter Open 5 2 5 Power connector with double Fuse and switch 5 2 6 Halogen light power adjustment 6 2 7 I O connector 6 3 Replacement of Deuterium or Halogen Lamp 7 4 Technical support 9 4 1 Getting...

Page 4: ...DUV version starts even lower at 175 nm for your deep UV experiments This version also offers twice the intensity at 200 nm The output of the AvaLight DH S is optimized for fibers or bundles upto 600...

Page 5: ...the SMA plug of the Light Source Connect your Fiber optic Cable first before starting the Deuterium Lamp Please be sure to avoid direct radiation of skin and eyes at the fiber output TTL shutter Shut...

Page 6: ...tching the Halogen switch downwards to ON the Halogen lamp switches on The Halogen LED indicates the operation mode Switching the Halogen switch upwards to OFF the Halogen lamp switches off Switch pos...

Page 7: ...he halogen power 2 7 I O connector For automatic shutter control to be controlled by software it is needed to connect an AvaSpec spectrometer to the AvaLight D H S with the interface cable IC DB26 2 N...

Page 8: ...he lamp envelope quartz glass should not be touched with bare fingers otherwise lifetime will be severely reduced Please follow these instructions for the bulb replacement 1 Switch off the light sourc...

Page 9: ...crews to secure the deuterium lamp 12 Remove the 2 screws of the halogen lamp fitting 13 Take out the halogen lamp fitting 14 Replace the halogen bulb with new 5W bulb caution do not touch the lamp gl...

Page 10: ...ches are ON before main power is switched on Put buttons for Deuterium and Halogen in OFF position before switching on main power 4 3 2 Deuterium Lamp Only FAULT POSSIBLE CAUSE REMEDY Power switch on...

Page 11: ...gen Lamp Only FAULT POSSIBLE CAUSE REMEDY Halogen Lamp does not work after switching on Halogen Lamp defective Replace Halogen Lamp LED does not light up after switching on Halogen lamp Internal Power...

Page 12: ...0 mm ca 5 kg total power for the specified wavelength range 5 1 Ordering Information AvaLight D S Deuterium light source 190 400 nm incl TTL shutter SR fibers needed AvaLight DH S Deuterium Halogen li...

Reviews: