Trouble Shooting 

 

4.3.1

 

Deuterium and Halogen lamps 

FAULT 

POSSIBLE CAUSE 

REMEDY 

Light doesn’t turn on, clicking 

sound is audible 

Front light switches are ON 

before main power is 

switched on. 

Put buttons for Deuterium and 

Halogen in OFF position before 

switching on main power. 

 

4.3.2

 

Deuterium Lamp Only 

FAULT 

POSSIBLE CAUSE 

REMEDY 

Power switch on, fan not 

working, nor one of the LED's 

or lights 

Line power not present  

Check line voltage  

Fuse defective 

Check /replace fuse(s) 

Deuterium Lamp does not 

ignite.  
The Two-Color-LED of 

deuterium lights up red 

(ERROR) 

Deuterium Lamp too hot 

Allow Deuterium Lamp to cool 

down.

 

 

Switch to OFF to reset the 

Deuterium lamp, then switch to 

ON to restart. 

Lifetime of Deuterium 

Lamp has expired 

Replace Deuterium Lamp 

Deuterium Lamp internal 

connection plug is not 

connected right 

Open Unit and connect connector 

plug 

Deuterium Lamp extinguishes 

during operation 

When this fault appears always switch off the deuterium power 

and then - after cool down - restart 

 
 

 

Summary of Contents for AvaLight-D-S

Page 1: ......

Page 2: ...oduct specifications and descriptions in this manual are subject to change without notice Avantes BV will not be responsible for errors and omissions in this manual or for direct or incidental damages...

Page 3: ...Shutter Open 5 2 5 Power connector with double Fuse and switch 5 2 6 Halogen light power adjustment 6 2 7 I O connector 6 3 Replacement of Deuterium or Halogen Lamp 7 4 Technical support 9 4 1 Getting...

Page 4: ...DUV version starts even lower at 175 nm for your deep UV experiments This version also offers twice the intensity at 200 nm The output of the AvaLight DH S is optimized for fibers or bundles upto 600...

Page 5: ...the SMA plug of the Light Source Connect your Fiber optic Cable first before starting the Deuterium Lamp Please be sure to avoid direct radiation of skin and eyes at the fiber output TTL shutter Shut...

Page 6: ...tching the Halogen switch downwards to ON the Halogen lamp switches on The Halogen LED indicates the operation mode Switching the Halogen switch upwards to OFF the Halogen lamp switches off Switch pos...

Page 7: ...he halogen power 2 7 I O connector For automatic shutter control to be controlled by software it is needed to connect an AvaSpec spectrometer to the AvaLight D H S with the interface cable IC DB26 2 N...

Page 8: ...he lamp envelope quartz glass should not be touched with bare fingers otherwise lifetime will be severely reduced Please follow these instructions for the bulb replacement 1 Switch off the light sourc...

Page 9: ...crews to secure the deuterium lamp 12 Remove the 2 screws of the halogen lamp fitting 13 Take out the halogen lamp fitting 14 Replace the halogen bulb with new 5W bulb caution do not touch the lamp gl...

Page 10: ...ches are ON before main power is switched on Put buttons for Deuterium and Halogen in OFF position before switching on main power 4 3 2 Deuterium Lamp Only FAULT POSSIBLE CAUSE REMEDY Power switch on...

Page 11: ...gen Lamp Only FAULT POSSIBLE CAUSE REMEDY Halogen Lamp does not work after switching on Halogen Lamp defective Replace Halogen Lamp LED does not light up after switching on Halogen lamp Internal Power...

Page 12: ...0 mm ca 5 kg total power for the specified wavelength range 5 1 Ordering Information AvaLight D S Deuterium light source 190 400 nm incl TTL shutter SR fibers needed AvaLight DH S Deuterium Halogen li...

Reviews: