Features and specifications
ThermoClamp-3 Controller
• Auto-tuning PID (Proportional Integral Derivative) temperature
control accurate to <1.0°C (often ±0.1˚C) with fuzzy logic
• Extremely low noise (analog heating) designed for electrophysiology
• Over-temperature protection
• Three types of user-configurable thermal runaway protection (patent
pending) with fuzzy logic constraints
• Smart voltage setpoint: digital, analog (BNC) or automatic
• Separate displays for both temperatures (control sensor and safety
sensor)
• Constant USB/serial data stream output
• Automatic heater failure detection and prevention
• Utilization of the safety sensor as secondary input for feedback control
Perfusion Pencil
®
Inline heater
• 1, 4, or 8 channels for solution heating with flow rates up to 5cc/min @
37˚C
• Temperature controlled bath perfusion inline heating with a flow rate
up to 5cc/min
• Small manifold volume for rapid switching
• Temperature can be maintained for the entire bath, or just for solutions
being delivered
• Perfusion Pencil tips are interchangeable / replaceable with standard
AutoMate tips
Electronic specifications
• Power supply: 100-240V AC, 50-60 Hz input; 12V, 5A output
• Heater output: 2 channels at quiet analog 0-12V up to 1A = 24 Watts
total
• Front panel BNC output voltage (to record the temperature): 40mV/
deg C
• Front panel input voltage (to control the setpoint): 20 deg C/V
• Serial baud rate: 230400
• Safety TC limit: 75˚C (adjustable)
• Max setpoint: 65˚C
• Default setpoint: 35˚C (adjustable)