6.4 Runaway detected @ divergence = xx.y
This error occurs when the controller detects that the bath sensor may
have fallen out, or the flow rate may have changed unexpectedly and
caused the bath sensor to cool rapidly. The sensitivity of this error
detection mode can be adjusted via the “ESB” parameter in the settings
menu.
6.5 Stopped flow detected @ divergence = xx.y
This error occurs when the controller detects that flow may be blocked
between the heater and the bath, or that the control sensor may be
slowly losing thermal contact with the bath fluid. The sensitivity of
this error detection mode can be adjusted via the “ESF” parameter in
the settings menu.
6.6 Control sensor is warming - no longer in error state.
This message will be displayed when the system recovers from an
error state because it detects the control sensor warming up again
after it is put back in the bath or flow resumes after previously being
stopped.
6.7 Low temp error - control temp too far below setpoint.
This message will be displayed with the control sensor temperature is
steadily falling below the setpoint and the “E LO” error is triggered.
23