Bath Perfusion / Water Jacket Feature
The Heated Perfusion Pencil includes an extra tube for heating a
flowing bath solution. This heated high-flow line will maintain the bath
temperature while quickly washing-out drugs applied through the tip.
The Pencil can maintain at least 37˚C with flow rates of 5cc/min.
This extra line can also be used with a water jacket on certain perfusion
chambers. A “water jacket” is usually a separate chamber or bath outside
a perfusion chamber for circulating a heating or cooling liquid. The
perfusion chamber must be equipped with a water jacket. It is not usually
something that can be added later. Simply provide a high flow of water
from a large reservoir through the Pencil, into the chamber’s water jacket,
and out to a drain or recirculating pump for ideal heating.
Figure 6.
Set up example of ThermoClamp temperature controller and
accessories.
Reservoirs
with the drug
solutions
Bath and Pencil TCs are
connected to their corresponding
sockets
Chamber heating
fluid source
Chamber with the cell
under examination
9